Pregnancy Hypertension 2018-Jan

Analytical validation of soluble fms-like tyrosine and placental growth factor assays on B·R·A·H·M·S KRYPTOR Compact Plus automated immunoassay platform.

Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Siaw Li Chan
Sarosh Rana
Sireesha Chinthala
Saira Salahuddin
Kiang-Teck J Yeo

Ключови думи

Резюме

BACKGROUND

Preeclampsia is one of the leading hypertensive disorders of pregnancy. Angiogenic biomarkers such as anti-angiogenic factor soluble fms-like tyrosine kinase 1 (sFlt1) and pro-angiogenic factor placental growth factor (PlGF) are involved in the pathophysiology of preeclampsia.

OBJECTIVE

The aim of this study is to validate the analytical performance of sFlt1 and PlGF on the B·R·A·H·M·S KRYPTOR Compact Plus (ThermoFisher Scientific).

METHODS

We examined K2-EDTA plasma samples from 50 patients on B·R·A·H·M·S KRYPTOR Compact Plus, an automated immunoassay platform. QC materials were used to assess intra- and inter-precision of the assay. Lower limit of quantitation and interference studies were determined using pooled patient plasma.

RESULTS

The sFlt1 and PlGF assays demonstrated an analytical measuring range of 90-69,000 pg/mL and 11-7000 pg/mL, respectively (r2 > 0.99). Lower limit of quantitation (20% CV) was interpolated to be 35 pg/mL for sFlt1 and 10 pg/mL for PlGF. Total precision for both assay displayed CVs of <10%. Interference studies showed that both assays were not significantly affected by hemolysis up to an H-index of 1100 for sFlt1 and 300 for PlGF; L- and I-index of 800 and 80 respectively for both assays. The Passing-Bablok regression analysis for sFlt1/PlGF yielded an equation of y = 1.05x + 0.02, and the Bland Altman analysis showed an average bias of 0.84.

CONCLUSIONS

Plasma levels of sFlt1 and PlGF measured on the B·R·A·H·M·S KRYPTOR Compact Plus platform demonstrate excellent analytical performance and are acceptable as clinical grade assays.

Присъединете се към нашата
страница във facebook

Най-пълната база данни за лечебни билки, подкрепена от науката

  • Работи на 55 езика
  • Билкови лекове, подкрепени от науката
  • Разпознаване на билки по изображение
  • Интерактивна GPS карта - маркирайте билките на място (очаквайте скоро)
  • Прочетете научни публикации, свързани с вашето търсене
  • Търсете лечебни билки по техните ефекти
  • Организирайте вашите интереси и бъдете в крак с научните статии, клиничните изследвания и патентите

Въведете симптом или болест и прочетете за билките, които биха могли да помогнат, напишете билка и вижте болестите и симптомите, срещу които се използва.
* Цялата информация се базира на публикувани научни изследвания

Google Play badgeApp Store badge