International Journal of Cardiology 2013-Sep

Left atrial appendage closure with the Amplatzer™ Cardiac Plug: impact of shape and device sizing on follow-up leaks.

Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Xavier Freixa
Apostolos Tzikas
Ayax Sobrino
Jason Chan
Arsène J Basmadjian
Réda Ibrahim

Ключови думи

Резюме

BACKGROUND

The Amplatzer Cardiac Plug (ACP) is a relatively new system for left atrial appendage (LAA) closure. In general, the particular design of the device and the variability of the LAA anatomy make the sizing process challenging. The objective of our study was to analyze the impact of the shape and sizing of the device on the presence of follow-up leaks.

METHODS

The shape was evaluated by angiography and classified as "strawberry-like", "square-like" and "tire-like". The presence of peri-device leaks was assessed by transesophageal echocardiography (TEE) and graded as: 1) severe, 2) major, 3) moderate, 4) minor and 5) absent.

RESULTS

Twenty five patients were included. Overall, the device was oversized by 3.38 ± 2.26 mm and 1.58 ± 2.75 mm in relation to the largest angiographic and TEE diameter respectively. None of the patients presented any intra-procedural complication. After discharge, one patient was admitted for cardiac tamponade and another for stroke. In both, the device was optimally deployed. Follow-up TEE was done after a mean time of 7.9 ± 5.9 months. The presence of a leak was associated with a lower degree of device over-sizing by TEE (0.06 ± 3.31 mm if leak vs. 2.73 ± 1.85 mm if no leak; p=0.04) and a trend towards by angiography (2.39 ± 3.06 mm if leak vs. 3.98 ± 1.57 mm if no leak; p=0.16). The shape of the device was not associated with the presence of TEE leaks. Neither the shape nor the presence of leaks was associated with worse clinical outcomes.

CONCLUSIONS

Although the shape and sizing of the device were not linked to any particular clinical outcome, we found an association between a lower degree of over-sizing and the presence of follow-up leaks.

Присъединете се към нашата
страница във facebook

Най-пълната база данни за лечебни билки, подкрепена от науката

  • Работи на 55 езика
  • Билкови лекове, подкрепени от науката
  • Разпознаване на билки по изображение
  • Интерактивна GPS карта - маркирайте билките на място (очаквайте скоро)
  • Прочетете научни публикации, свързани с вашето търсене
  • Търсете лечебни билки по техните ефекти
  • Организирайте вашите интереси и бъдете в крак с научните статии, клиничните изследвания и патентите

Въведете симптом или болест и прочетете за билките, които биха могли да помогнат, напишете билка и вижте болестите и симптомите, срещу които се използва.
* Цялата информация се базира на публикувани научни изследвания

Google Play badgeApp Store badge