Circulation: Cardiovascular Interventions 2017-Jun

Outcome With the Repositionable and Retrievable Boston Scientific Lotus Valve Compared With the Balloon-Expandable Edwards Sapien 3 Valve in Patients Undergoing Transfemoral Aortic Valve Replacement.

Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Julia Seeger
Birgid Gonska
Wolfgang Rottbauer
Jochen Wöhrle

Ключови думи

Резюме

BACKGROUND

New generation devices for transfemoral aortic valve replacement were optimized on valve positioning and reduction of residual aortic regurgitation. We compared 30-day, 12-month, and 24-month outcomes of the Boston Scientific Lotus valve (Lotus) and the balloon-expandable Edwards Sapien 3 (ES3) valve. Primary end point was all-cause mortality or disabling stroke within 12 months.

RESULTS

Between 2014 and 2016, 537 patients were enrolled at our center, and 202 patients received Lotus and 335 ES3. There was no residual moderate or severe aortic regurgitation. Rate of mild aortic regurgitation was lower with the repositionable and retrievable Lotus valve compared with the ES3. Rate of pacemaker implantation was significantly higher with the Lotus valve compared with the ES3 valve (36.1% versus 14.9%, P<0.01). Valve Academic Research Consortium-2 early safety end point at 30 days was 7.4% with both devices with no difference in all-cause mortality (Lotus, 1.9%; ES3, 1.8%; P=0.87), rate of disabling stroke (Lotus, 1.5%; ES3, 2.1%; P=0.62), or major vascular complications (Lotus, 2.9%; ES3, 2.4%; P=0.69). The primary end point at 12 months was similar between groups. In a propensity score-matched comparison, there was no difference in the primary end point within 12 months (Lotus, 15.5%; ES3, 18.6%; P=0.69) and 24 months (Lotus, 21.9%; ES3, 26.4%; P=0.49).

CONCLUSIONS

Transfemoral aortic valve replacement with the ES3 and the Lotus were associated with similar 30-day, 12-month, and 24-month clinical outcomes. Need for permanent pacemaker implantation was significantly higher with the repositionable Lotus device.

BACKGROUND

URL: http://www.clinicaltrials.gov. Unique identifier: NCT02162069.

Присъединете се към нашата
страница във facebook

Най-пълната база данни за лечебни билки, подкрепена от науката

  • Работи на 55 езика
  • Билкови лекове, подкрепени от науката
  • Разпознаване на билки по изображение
  • Интерактивна GPS карта - маркирайте билките на място (очаквайте скоро)
  • Прочетете научни публикации, свързани с вашето търсене
  • Търсете лечебни билки по техните ефекти
  • Организирайте вашите интереси и бъдете в крак с научните статии, клиничните изследвания и патентите

Въведете симптом или болест и прочетете за билките, които биха могли да помогнат, напишете билка и вижте болестите и симптомите, срещу които се използва.
* Цялата информация се базира на публикувани научни изследвания

Google Play badgeApp Store badge