Neonatology 2017

Reliability of Pulse Oximetry during Progressive Hypoxia, Cardiopulmonary Resuscitation, and Recovery in a Piglet Model of Neonatal Hypoxic Cardiac Arrest.

Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Mohammad Ahmad Hassan
Claudia Weber
Markus Waitz
Li Huang
Helmut D Hummler
Marc Robin Mendler

Ключови думи

Резюме

BACKGROUND

Pulse oximetry is widely used in intensive care and emergency conditions to monitor arterial oxygenation and to guide oxygen therapy.

OBJECTIVE

To study the reliability of pulse oximetry in comparison with CO-oximetry in newborn piglets during progressive hypoxia, cardiac arrest, cardiopulmonary resuscitation (CPR), and after return of spontaneous circulation (ROSC).

METHODS

Thirty-three newborn piglets were exposed to hypoxia until asystole occurred and then resuscitated until ROSC. Arterial oxygen saturation was monitored continuously by pulse oximetry (SpO2) with one sensor applied to the wrist of the right forelimb (FL) and another to the thigh of the left hind limb (HL). Arterial functional oxygen saturation (SaO2) was measured at baseline and at predefined intervals during each phase of the experiment. SpO2 was compared with coinciding SaO2 values and bias considered whenever the difference (SpO2 - SaO2) was beyond ±5%.

RESULTS

Bias values were lower at the baseline measurements (-3.7 ± 2.3% in FL and -4.1 ± 3.4% in HL) as well as after ROSC (1.5 ± 4.2% in FL and 0.2 ± 4.6% in HL) with higher precision and accuracy than during other experiment phases. During hypoxia induction, cardiac arrest, and CPR, there was a marked decrease in precision and accuracy as well as an increase in bias up to 43 ± 26 and 56 ± 27% in FL and HL, respectively, over a range of SaO2 from 13 to 51%.

CONCLUSIONS

Pulse oximetry showed increased bias and decreased accuracy and precision during marked hypoxemia in a model of neonatal hypoxic cardiac arrest.

Присъединете се към нашата
страница във facebook

Най-пълната база данни за лечебни билки, подкрепена от науката

  • Работи на 55 езика
  • Билкови лекове, подкрепени от науката
  • Разпознаване на билки по изображение
  • Интерактивна GPS карта - маркирайте билките на място (очаквайте скоро)
  • Прочетете научни публикации, свързани с вашето търсене
  • Търсете лечебни билки по техните ефекти
  • Организирайте вашите интереси и бъдете в крак с научните статии, клиничните изследвания и патентите

Въведете симптом или болест и прочетете за билките, които биха могли да помогнат, напишете билка и вижте болестите и симптомите, срещу които се използва.
* Цялата информация се базира на публикувани научни изследвания

Google Play badgeApp Store badge