International Journal of Cardiology 2013-Sep

Left atrial appendage closure with the Amplatzer™ Cardiac Plug: impact of shape and device sizing on follow-up leaks.

Samo registrirani korisnici mogu prevoditi članke
Prijavite se / prijavite se
Veza se sprema u međuspremnik
Xavier Freixa
Apostolos Tzikas
Ayax Sobrino
Jason Chan
Arsène J Basmadjian
Réda Ibrahim

Ključne riječi

Sažetak

BACKGROUND

The Amplatzer Cardiac Plug (ACP) is a relatively new system for left atrial appendage (LAA) closure. In general, the particular design of the device and the variability of the LAA anatomy make the sizing process challenging. The objective of our study was to analyze the impact of the shape and sizing of the device on the presence of follow-up leaks.

METHODS

The shape was evaluated by angiography and classified as "strawberry-like", "square-like" and "tire-like". The presence of peri-device leaks was assessed by transesophageal echocardiography (TEE) and graded as: 1) severe, 2) major, 3) moderate, 4) minor and 5) absent.

RESULTS

Twenty five patients were included. Overall, the device was oversized by 3.38 ± 2.26 mm and 1.58 ± 2.75 mm in relation to the largest angiographic and TEE diameter respectively. None of the patients presented any intra-procedural complication. After discharge, one patient was admitted for cardiac tamponade and another for stroke. In both, the device was optimally deployed. Follow-up TEE was done after a mean time of 7.9 ± 5.9 months. The presence of a leak was associated with a lower degree of device over-sizing by TEE (0.06 ± 3.31 mm if leak vs. 2.73 ± 1.85 mm if no leak; p=0.04) and a trend towards by angiography (2.39 ± 3.06 mm if leak vs. 3.98 ± 1.57 mm if no leak; p=0.16). The shape of the device was not associated with the presence of TEE leaks. Neither the shape nor the presence of leaks was associated with worse clinical outcomes.

CONCLUSIONS

Although the shape and sizing of the device were not linked to any particular clinical outcome, we found an association between a lower degree of over-sizing and the presence of follow-up leaks.

Pridružite se našoj
facebook stranici

Najkompletnija baza ljekovitog bilja potpomognuta naukom

  • Radi na 55 jezika
  • Biljni lijekovi potpomognuti naukom
  • Prepoznavanje biljaka po slici
  • Interaktivna GPS karta - označite bilje na lokaciji (uskoro)
  • Pročitajte naučne publikacije povezane sa vašom pretragom
  • Pretražite ljekovito bilje po učincima
  • Organizirajte svoja interesovanja i budite u toku sa istraživanjem vijesti, kliničkim ispitivanjima i patentima

Upišite simptom ili bolest i pročitajte o biljkama koje bi mogle pomoći, unesite travu i pogledajte bolesti i simptome protiv kojih se koristi.
* Sve informacije temelje se na objavljenim naučnim istraživanjima

Google Play badgeApp Store badge