American Journal of Roentgenology 2011-Jun

Decreased detection of hypovascular liver tumors with MDCT in obese patients: a phantom study.

Kun registrerede brugere kan oversætte artikler
Log ind / Tilmeld
Linket gemmes på udklipsholderen
Sebastian T Schindera
Jaled Charimo Torrente
Thomas D Ruder
Hanno Hoppe
Daniele Marin
Rendon C Nelson
Zsolt Szucs-Farkas

Nøgleord

Abstrakt

OBJECTIVE

The purpose of this article is to assess the impact of large patient size on the detection of hypovascular liver tumors with MDCT and the effect of a noise filter on image quality and lesion detection in obese patients.

METHODS

A liver phantom with 45 hypovascular tumors (diameters of 5, 10, and 15 mm) was placed into two water containers mimicking intermediate and large patients. The containers were scanned with a 64-MDCT scanner. The CT dataset from the large phantom was postprocessed using a noise filter. The image noise was measured and the contrast-to-noise ratio (CNR) of the tumors was calculated. Tumor detection was independently performed by three radiologists in a blinded fashion.

RESULTS

The application of the noise filter in the large phantom yielded a reduction of image noise by 42% (p < 0.0001). The CNR values of the tumors in the nonfiltered and filtered large phantom were lower than that in the intermediate phantom (p < 0.05). In the non-filtered and filtered large phantom, 25% and 19% fewer tumors, respectively, were detected on average compared with the intermediate phantom (p < 0.01).

CONCLUSIONS

The risk of missing hypovascular liver tumors with CT is substantially increased in large patients. A noise filter improves image quality in obese patients.

Deltag i vores
facebook-side

Den mest komplette database med medicinske urter understøttet af videnskab

  • Arbejder på 55 sprog
  • Urtekurer, der understøttes af videnskab
  • Urtegenkendelse ved billede
  • Interaktivt GPS-kort - tag urter på stedet (kommer snart)
  • Læs videnskabelige publikationer relateret til din søgning
  • Søg medicinske urter efter deres virkninger
  • Organiser dine interesser og hold dig opdateret med nyhedsundersøgelser, kliniske forsøg og patenter

Skriv et symptom eller en sygdom, og læs om urter, der kan hjælpe, skriv en urt og se sygdomme og symptomer, den bruges mod.
* Al information er baseret på offentliggjort videnskabelig forskning

Google Play badgeApp Store badge