Journal of Prosthetic Dentistry 2005-Nov

In vitro wear resistance of three types of composite resin denture teeth.

Kun registrerede brugere kan oversætte artikler
Log ind / Tilmeld
Linket gemmes på udklipsholderen
Jianyu Zeng
Yohei Sato
Chikahiro Ohkubo
Toshio Hosoi

Nøgleord

Abstrakt

BACKGROUND

The wear resistance of artificial teeth is important to the life of the removable prosthesis for the edentulous patient. The ability of artificial teeth to maintain a stable occlusal relationship over time may be influenced by this property.

OBJECTIVE

The purpose of this in vitro study was to evaluate the wear resistance of artificial teeth based on composite resin filler content when opposed by a denture tooth antagonist.

METHODS

The right first molars (both maxillary and mandibular) of 3 composite resin denture teeth (Endura, 47% organic filler; Duradent, 42% organic and inorganic filler; and Duracross, 68% organic filler) were secured in an in vitro 2-body wear-testing apparatus that simulated masticatory function (60 strokes/min, grinding distance of 2 mm, under 37 degrees C flowing water). Wear resistance was assessed as volume loss (mm3) and height loss (mm) at 5 kgf (grinding force) after 50,000 strokes (6 pairs for each type). The Vickers microhardness (VHN) of the occlusal surfaces was measured with a load of 50 g and a loading time of 30 seconds. After testing, the worn surfaces were observed with a scanning electron microscope (SEM). Mean values were analyzed using a 1-way analysis of variance followed by the Fisher test (alpha=.05).

RESULTS

The volume loss of Duradent teeth (0.87 +/- 0.03 mm3) was significantly higher than of the 2 other types of teeth tested (P<.001). The Duracross teeth (0.22 +/- 0.08 mm3 and 0.11 +/- 0.04 mm) exhibited the best wear resistance of the denture teeth tested. The wear resistance of the materials tested was consistent with their hardness (Endura 27.54 +/- 0.67 VHN, Duradent 24.00 +/- 0.60 VHN, Duracross 39.64 +/- 3.11 VHN). SEM observation showed that many filler particles were missing from the matrix of Duradent.

CONCLUSIONS

The wear resistance among the 3 types of teeth was significantly different; Duracross and Endura were more wear resistant than Duradent. Although the densities as reported by the manufacturer were similar, hardness and microstructure were the primary factors affecting the wear resistance of artificial teeth. Hardness appears to correlate with wear resistance since the wear resistance of Duradent with a lower VHN of 24 was greater than the other 2 types of teeth with greater hardness.

Deltag i vores
facebook-side

Den mest komplette database med medicinske urter understøttet af videnskab

  • Arbejder på 55 sprog
  • Urtekurer, der understøttes af videnskab
  • Urtegenkendelse ved billede
  • Interaktivt GPS-kort - tag urter på stedet (kommer snart)
  • Læs videnskabelige publikationer relateret til din søgning
  • Søg medicinske urter efter deres virkninger
  • Organiser dine interesser og hold dig opdateret med nyhedsundersøgelser, kliniske forsøg og patenter

Skriv et symptom eller en sygdom, og læs om urter, der kan hjælpe, skriv en urt og se sygdomme og symptomer, den bruges mod.
* Al information er baseret på offentliggjort videnskabelig forskning

Google Play badgeApp Store badge