Journal of Veterinary Emergency and Critical Care 2016-Sep

Validation of the Accutrend Plus point-of-care triglyceride analyzer in horses, ponies, and donkeys.

Kun registrerede brugere kan oversætte artikler
Log ind / Tilmeld
Linket gemmes på udklipsholderen
Michaela C Pongratz
Hannah K Junge
Barbara Riond
Colin C Schwarzwald

Nøgleord

Abstrakt

OBJECTIVE

To assess the accuracy and reliability of a point-of-care (POC) triglyceride analyzer and to establish reference intervals for blood ([TRIG]POC/WB ) and plasma triglyceride concentrations ([TRIG]POC/PL ) in horses, ponies, and donkeys.

METHODS

Prospective study.

METHODS

University teaching hospital.

METHODS

120 adult healthy equids (78 horses and ponies, 42 donkeys) and 79 equids suffering from hypertriglyceridemia (73 horses and ponies, 6 donkeys).

METHODS

None.

RESULTS

[TRIG]POC/WB and [TRIG]POC/PL were measured using a POC analyzer and plasma triglyceride concentrations were measured using a standard laboratory assay ([TRIG]LAB/PL ). Reference intervals were determined. Test accuracy was assessed by Bland-Altman comparison of POC measurements with standard laboratory measurements and by evaluating linearity of dilutional series. Test reliability was assessed by repeated serial measurements. [TRIG]POC/WB and [TRIG]POC/PL were below the analytic range of the POC assay (<0.8 mmol/L [<70 mg/dL]) in healthy horses and ponies, whereas the reference intervals were 0.82-3.14 mmol/L (73-278 mg/dL) and 0.87-3.02 mmol/L (77-267 mg/dL), respectively, in donkeys. The POC analyzer systematically overestimated triglyceride concentrations when compared to a standard laboratory assay. The difference between [TRIG]POC/WB and [TRIG]POC/PL was small and clinically negligible. The coefficient of variation of repeated measures performed on the POC analyzer was below 10% for [TRIG]POC/WB and [TRIG]POC/PL , both in horses and donkeys and at all concentration ranges.

CONCLUSIONS

The POC analyzer allows accurate and reliable measurement of [TRIG]POC/WB and [TRIG]POC/PL in horses, ponies, and donkeys in clinical settings. Assay-specific reference intervals should be determined for diagnosis and clinical monitoring of hypertriglyceridemia in equids.

Deltag i vores
facebook-side

Den mest komplette database med medicinske urter understøttet af videnskab

  • Arbejder på 55 sprog
  • Urtekurer, der understøttes af videnskab
  • Urtegenkendelse ved billede
  • Interaktivt GPS-kort - tag urter på stedet (kommer snart)
  • Læs videnskabelige publikationer relateret til din søgning
  • Søg medicinske urter efter deres virkninger
  • Organiser dine interesser og hold dig opdateret med nyhedsundersøgelser, kliniske forsøg og patenter

Skriv et symptom eller en sygdom, og læs om urter, der kan hjælpe, skriv en urt og se sygdomme og symptomer, den bruges mod.
* Al information er baseret på offentliggjort videnskabelig forskning

Google Play badgeApp Store badge