Clinical Chemistry 2008-Mar

Active B12: a rapid, automated assay for holotranscobalamin on the Abbott AxSYM analyzer.

Nur registrierte Benutzer können Artikel übersetzen
Einloggen Anmelden
Der Link wird in der Zwischenablage gespeichert
Jeff Brady
Lesley Wilson
Lynda McGregor
Edward Valente
Lars Orning

Schlüsselwörter

Abstrakt

BACKGROUND

Conventional tests for vitamin B(12) deficiency measure total serum vitamin B12, whereas only that portion of vitamin B12 carried by transcobalamin (holotranscobalamin) is metabolically active. Measurement of holotranscobalamin (holoTC) may be more diagnostically accurate for detecting B(12) deficiency that requires therapy. We developed an automated assay for holoTC that can be used on the Abbott AxSYM immunoassay analyzer.

METHODS

AxSYM Active B12 is a 2-step sandwich microparticle enzyme immunoassay. In step 1, a holoTC-specific antibody immobilized onto latex microparticles captures holoTC in samples of serum or plasma. In step 2, the captured holoTC is detected with a conjugate of alkaline phosphatase and antiTC antibody.

RESULTS

Neither apoTC nor haptocorrin exhibited detectable cross-reactivity. The detection limit was < or = 0.1 pmol/L. Within-run and total imprecision (CV ranges) were 3.4%-5.1% and 6.3%-8.5%, respectively. Assay CVs were < 20% from at least 3 pmol/L to 107 pmol/L. With diluted serum samples, measured concentrations were 104%-114% of the expected values in the working range of the assay. No interference from bilirubin, hemoglobin, triglycerides, erythrocytes, rheumatoid factor, or total protein was detected at expected (abnormal) concentrations. A comparison of the AxSYM Active B12 assay with a commercial RIA for holoTC yielded the regression equation: AxSYM = 0.98RIA + 4.7 pmol/L (S(y x), 11.4 pmol/L; n = 204). Assay throughput was 45 tests/h. A 95% reference interval of 19-134 pmol/L holoTC was established with samples from 292 healthy individuals.

CONCLUSIONS

The AxSYM Active B12 assay allows rapid, precise, sensitive, specific, and automated measurement of human holoTC in serum and plasma.

Treten Sie unserer
Facebook-Seite bei

Die vollständigste Datenbank für Heilkräuter, die von der Wissenschaft unterstützt wird

  • Arbeitet in 55 Sprachen
  • Von der Wissenschaft unterstützte Kräuterkuren
  • Kräutererkennung durch Bild
  • Interaktive GPS-Karte - Kräuter vor Ort markieren (in Kürze)
  • Lesen Sie wissenschaftliche Veröffentlichungen zu Ihrer Suche
  • Suchen Sie nach Heilkräutern nach ihrer Wirkung
  • Organisieren Sie Ihre Interessen und bleiben Sie über Neuigkeiten, klinische Studien und Patente auf dem Laufenden

Geben Sie ein Symptom oder eine Krankheit ein und lesen Sie über Kräuter, die helfen könnten, geben Sie ein Kraut ein und sehen Sie Krankheiten und Symptome, gegen die es angewendet wird.
* Alle Informationen basieren auf veröffentlichten wissenschaftlichen Forschungsergebnissen

Google Play badgeApp Store badge