Therapeutic Drug Monitoring 2014-Apr

Fast and highly selective LC-MS/MS screening for THC and 16 other abused drugs and metabolites in human hair to monitor patients for drug abuse.

Only registered users can translate articles
Log In/Sign up
The link is saved to the clipboard
Remco A Koster
Jan-Willem C Alffenaar
Ben Greijdanus
Joanneke E L VanDernagel
Donald R A Uges

Keywords

Abstract

BACKGROUND

To facilitate the monitoring of drug abuse by patients, a method was developed and validated for the analysis of amphetamine, methamphetamine, 3,4-methylenedioxymethamphetamine, methylenedioxyamphetamine, methylenedioxyethylamphetamine, methylphenidate, cocaine, benzoylecgonine, morphine, codeine, heroin, 6-monoacteylmorphine, methadone, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), delta-9-tetrahydrocannabinol (THC), nicotine, and cotinine in human hair.

METHODS

The hair preparation method contains a 3-step wash procedure with dichloromethane followed by a simultaneous hair pulverization and extraction procedure with disposable metal balls. The developed liquid chromatography tandem mass spectrometry method uses a single injection to detect and confirm all 17 abused drugs, including THC, within 4.8 minutes.

RESULTS

Nicotine was validated with a linear range of 800-25,000 pg/mg hair, and all other substances were validated with a linear range of 30.0-2500 pg/mg hair. For inaccuracy and imprecision, the overall bias did not exceed -8.2% and the overall coefficient of variation did not exceed 17.7%. Autosampler stability was proven for 48 hours at 10°C for all substances. Analytical cutoff concentrations were defined for each substance at the lowest validated inaccuracy and imprecision concentration with a bias and coefficient of variation within 15% and qualifier/quantifier ratios within 20% of the set ratio. The analytical cutoff concentrations were 200 pg/mg for codeine and 80.0 pg/mg for 6-MAM, heroin, EDDP, and THC. The analytical cutoff concentration for nicotine was 800 pg/mg and for all other validated substances 30.0 pg/mg. This method was successfully applied to analyze hair samples from patients who were monitored for drug abuse. Hair samples of 47 subjects (segmented into 129 samples) showed 3,4-methylenedioxymethamphetamine, methylphenidate, cocaine, benzoylecgonine, codeine, methadone, EDDP, THC, nicotine, and cotinine above the analytical cutoff.

CONCLUSIONS

The method was fully validated, including the validation of the qualifier/quantifier ratios. The analysis of real hair samples proved the efficacy of the developed method for monitoring drug abuse. The results obtained by this method provide the physician or health-care professional with extensive information about actual drug abuse or relapse and can be used for patient-specific therapy.

Join our facebook page

The most complete medicinal herbs database backed by science

  • Works in 55 languages
  • Herbal cures backed by science
  • Herbs recognition by image
  • Interactive GPS map - tag herbs on location (coming soon)
  • Read scientific publications related to your search
  • Search medicinal herbs by their effects
  • Organize your interests and stay up do date with the news research, clinical trials and patents

Type a symptom or a disease and read about herbs that might help, type a herb and see diseases and symptoms it is used against.
*All information is based on published scientific research

Google Play badgeApp Store badge