European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies 1992-Sep

A multicentre evaluation of the Ektachem DT60-, Reflotron- and Seralyzer III systems.

Solo los usuarios registrados pueden traducir artículos
Iniciar sesión Registrarse
El enlace se guarda en el portapapeles.
M Römer
R Haeckel
A Henco
M Vogt
L Thomas
H E Keller
W Appel

Palabras clave

Abstracto

The analytical performance of the three analytical systems Reflotron, Ektachem DT60 and Seralyzer III was studied according to ECCLS guidelines (1) and partly according to a protocol of the Société Francaise de Biologie Clinique (SFBC) (2) in a multicentre evaluation involving four laboratories. The determination of 11 analytes led to more than 180,000 data. With the Ektachem DT60, imprecision was acceptable for all analytes except for sodium. With the Reflotron, imprecision for glucose and creatinine was wider than the acceptable limits. With the Seralyzer III, imprecision for glucose, uric acid, cholesterol, creatinine and potassium was not within acceptance limits. The recovery of system assigned control sera values was acceptable for all analytes, except for glucose, creatinine and sodium with the Ektachem DT60 system, and for all but glucose and cholesterol with the Reflotron. On the Seralyzer III, the limits of acceptance were exceeded only with the creatinine assay. The recovery of the reference method values with Kontrollogen L caused problems with all three systems and for all analytes. Only 30% of the mean values of Kontrollogen L measured with the Ektachem and with the Seralyzer III were within the limits of acceptance. 40% of the mean values determined with the Reflotron were inside these limits. The upper limits of linearity as claimed by the manufacturers were obtained with all analytes and systems with the exception of cholesterol on all systems and the creatinine assay on the Seralyzer III. The systems under test and several different comparison methods showed good agreement for the analysis of patient samples, except in one laboratory for the analysis of sodium and aspartate aminotransferase with the Ektachem system and potassium with the Seralyzer III. Turbidity showed no significant influence on the measurements of all analytes and all systems. Haemolysis, hyperproteinaemia, and bilirubinaemia affected several methods on all three systems. A start time delay of up to 60 s did not affect the results of the Reflotron, except in the case of the triacylglycerol assay, which was affected by start time delays greater than 45 s. The results of 4 assays on the Seralyzer III were decreased considerably by a delayed start time (triacylglycerols and creatinine above 5 s, aspartate aminotransferase above 35 s and creatine kinase above 15 s). For reliable results from all the assays in each of the three analytical systems, it was necessary to use the prescribed sample volume within certain limits. The practicability of all analytical systems tested was found to be very good. A field study was conducted with the Reflotron system. The analyte concentration was determined in venous blood from various patients. In 25 out of 30 experiments, the results of the "field" laboratories showed a greater spread about the fitting line than those obtained in the "expert" laboratories.

Únete a nuestra
página de facebook

La base de datos de hierbas medicinales más completa respaldada por la ciencia

  • Funciona en 55 idiomas
  • Curas a base de hierbas respaldadas por la ciencia
  • Reconocimiento de hierbas por imagen
  • Mapa GPS interactivo: etiquete hierbas en la ubicación (próximamente)
  • Leer publicaciones científicas relacionadas con su búsqueda
  • Buscar hierbas medicinales por sus efectos.
  • Organice sus intereses y manténgase al día con las noticias de investigación, ensayos clínicos y patentes.

Escriba un síntoma o una enfermedad y lea acerca de las hierbas que podrían ayudar, escriba una hierba y vea las enfermedades y los síntomas contra los que se usa.
* Toda la información se basa en investigaciones científicas publicadas.

Google Play badgeApp Store badge