Telemedicine Journal and e-Health 2017-Oct

Accuracy and Reliability of a Handheld, Nonmydriatic Fundus Camera for the Remote Detection of Optic Disc Edema.

Solo los usuarios registrados pueden traducir artículos
Iniciar sesión Registrarse
El enlace se guarda en el portapapeles.
Lulu Bursztyn
Maria A Woodward
Wayne T Cornblath
Hilary M Grabe
Jonathan D Trobe
Leslie Niziol
Lindsey B De Lott

Palabras clave

Abstracto

Optic disc edema can be an important indicator of serious neurological disease, but is poorly detected using the direct ophthalmoscope. Portable fundus photography may overcome this difficulty.

The purpose of this study was to determine the sensitivity and specificity of a handheld, nonmydriatic fundus camera for the detection of optic disc edema.

Retrospective review of nonmydriatic optic disc photographs taken with a portable fundus camera (Pictor Plus; Volk Optical, Mentor, OH) from the University of Michigan Neuro-Ophthalmology Clinics. We included 103 consecutive eyes with optic disc edema and 103 consecutive eyes without optic disc edema of 109 patients. Four masked neuro-ophthalmologists graded a single photograph of each optic disc presented in randomized order and documented the presence of optic disc edema. Sensitivity and specificity of graders' photographic interpretation was compared with clinical examinations. Reliability of assessments within and between graders was determined using kappa statistics.

The sensitivity and specificity for detection of optic disc edema were 71.8-92.2% and 81.6-95.2%, respectively. Photos were found to be ungradable in 0-8.3% of cases. The intergrader reliabilities ranged from 0.60 [95% confidence interval (CI): 0.52-0.67] to 0.72 (95% CI: 0.66-0.77). Intragrader reliability ranged from 0.76 (95% CI: 0.63-0.92) to 0.82 (95% CI: 0.69-0.95).

Photographs taken with portable, nonmydriatic technology met threshold sensitivity and specificity for remote screening for optic disc edema when performed by most, but not all graders. Reliability between graders was moderate-strong and strong within individual providers.

Portable photography holds promise for use in remote screening of optic disc edema.

Únete a nuestra
página de facebook

La base de datos de hierbas medicinales más completa respaldada por la ciencia

  • Funciona en 55 idiomas
  • Curas a base de hierbas respaldadas por la ciencia
  • Reconocimiento de hierbas por imagen
  • Mapa GPS interactivo: etiquete hierbas en la ubicación (próximamente)
  • Leer publicaciones científicas relacionadas con su búsqueda
  • Buscar hierbas medicinales por sus efectos.
  • Organice sus intereses y manténgase al día con las noticias de investigación, ensayos clínicos y patentes.

Escriba un síntoma o una enfermedad y lea acerca de las hierbas que podrían ayudar, escriba una hierba y vea las enfermedades y los síntomas contra los que se usa.
* Toda la información se basa en investigaciones científicas publicadas.

Google Play badgeApp Store badge