Spanish
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
CardioVascular and Interventional Radiology

Hyperthermia in the treatment of cholangiocarcinoma: development and testing of an endobiliary microwave device.

Solo los usuarios registrados pueden traducir artículos
Iniciar sesión Registrarse
El enlace se guarda en el portapapeles.
Sara Mantero
Iginio Longo
Gianfranco Beniamino Fiore
Aldo Severini

Palabras clave

Abstracto

OBJECTIVE

The aim of this study was to create, perfect and test a hyperthermia balloon catheter for local treatment of cholangiocarcinoma. The device should induce hyperthermia in tumor tissue by acting locally in the bile duct lumen in contact with the tumor-infiltrated duct wall. In addition, it should exert tissue compression to cause an appreciable reduction in tumor microvasculature flow, thus improving thermal performance.

METHODS

According to the design specifications, the working temperature range should allow the device to be used for hyperthermia therapy either in combination with radiation and/or chemotherapy (radio/chemo-sensitizing effect at 42-45 degrees C) or alone (induction of coagulation necrosis above 65 degrees C). The balloon serves as an anchoring system during treatment and as a functional element to induce tissue compression. In vitro mechanical evaluation of the pressure/volume relationship, with the balloon inside rigid walled conduits, was performed. The heating shape around the catheter tip was determined by egg-white heating tests (coagulation at about 65 degrees C). Moreover, heating tests were carried out with explanted pig liver parenchyma. The temperature profile over time at different depths from the catheter axis was traced.

RESULTS

Three prototypes were manufactured. Mechanical functional tests showed that a 14 Fr deflated diameter balloon was suitable for bile duct diameters up to 11 mm. Thermal egg-white tests produced 4 cm long, 3 cm in diameter ellipsoidal heating figures in 30 min. In the biological tissue tests a coagulated area of similar geometry and comparable volume was produced.

CONCLUSIONS

The results of tests confirm the device's usefulness and versatility.

Únete a nuestra
página de facebook

La base de datos de hierbas medicinales más completa respaldada por la ciencia

  • Funciona en 55 idiomas
  • Curas a base de hierbas respaldadas por la ciencia
  • Reconocimiento de hierbas por imagen
  • Mapa GPS interactivo: etiquete hierbas en la ubicación (próximamente)
  • Leer publicaciones científicas relacionadas con su búsqueda
  • Buscar hierbas medicinales por sus efectos.
  • Organice sus intereses y manténgase al día con las noticias de investigación, ensayos clínicos y patentes.

Escriba un síntoma o una enfermedad y lea acerca de las hierbas que podrían ayudar, escriba una hierba y vea las enfermedades y los síntomas contra los que se usa.
* Toda la información se basa en investigaciones científicas publicadas.

Google Play badgeApp Store badge