International Orthopaedics 2015-Jun

Minimum five-year wear rate of metal-on-highly cross-linked polyethylene in primary total hip arthroplasty.

Solo los usuarios registrados pueden traducir artículos
Iniciar sesión Registrarse
El enlace se guarda en el portapapeles.
Caroline Scemama
Claudio Dora
Jean Langlois
Moussa Hamadouche

Palabras clave

Abstracto

OBJECTIVE

Highly cross-linked polyethylenes (PE) have been developed with encouraging results in terms of wear. Another body of the literature has indicated potential catastrophic failures related to reduced fatigue properties and oxidation. Each PE available on the market has its own processing characteristics. The aim of this retrospective study was to evaluate the minimum five-year wear properties of an original highly cross-linked PE in a consecutive series of primary THAs.

METHODS

Between August 2005 and December 2007, 80 patients with a mean age of 62.7 years were included. All patients had a 28-mm CoCr femoral head articulating with a highly cross-linked insert (Highcross®, Medacta SA) that was 100 Mrads gamma radiated, remelted at 150 °C, and ethylene oxide sterilized. The primary criterion for evaluation was the femoral head penetration, as measured by Hip Analysis Suite software. The steady state wear was also calculated. Functional results were evaluated according to the WOMAC score.

RESULTS

Complete data were available for analysis in 67 patients at a mean follow-up of 5.5 years. The mean femoral head penetration was 0.128 ± 0.62 mm and the steady state wear was-0.025 ± 0.22 mm/year. The WOMAC score significantly decreased from 16.5 ± 5.93 pre-operatively to 4.12 ± 5.5 at the latest follow-up (p <0.001).

CONCLUSIONS

The minimal five-year results of this retrospective study indicate that this particular highly cross-linked and remelted polyethylene had a low wear rate. Longer-term results are needed to warrant that these mid-term data will generate less osteolysis and resultant aseptic loosening.

Únete a nuestra
página de facebook

La base de datos de hierbas medicinales más completa respaldada por la ciencia

  • Funciona en 55 idiomas
  • Curas a base de hierbas respaldadas por la ciencia
  • Reconocimiento de hierbas por imagen
  • Mapa GPS interactivo: etiquete hierbas en la ubicación (próximamente)
  • Leer publicaciones científicas relacionadas con su búsqueda
  • Buscar hierbas medicinales por sus efectos.
  • Organice sus intereses y manténgase al día con las noticias de investigación, ensayos clínicos y patentes.

Escriba un síntoma o una enfermedad y lea acerca de las hierbas que podrían ayudar, escriba una hierba y vea las enfermedades y los síntomas contra los que se usa.
* Toda la información se basa en investigaciones científicas publicadas.

Google Play badgeApp Store badge