Resuscitation 2006-May

A new device producing manual sternal compression with thoracic constraint for cardiopulmonary resuscitation.

فقط کاربران ثبت نام شده می توانند مقالات را ترجمه کنند
ورود به سیستم / ثبت نام
پیوند در کلیپ بورد ذخیره می شود
James T Niemann
John P Rosborough
Leo Kassabian
Bobak Salami

کلید واژه ها

خلاصه

OBJECTIVE

Blood flow during conventional cardiopulmonary resuscitation (CPR) is usually less than adequate to sustain vital organ perfusion. A new chest compression device (LifeBelt) which compresses both the sternum and the lateral thoraces (compression and thoracic constraint) has been developed. The device is light weight, portable, manually powered and mechanically advantaged to minimize user fatigue. The purpose of this study was to evaluate the mechanism of blood flow with the device, determine the optimal compression force and compare the device to standard manual CPR in a swine arrest model.

METHODS

Following anesthesia and instrumentation, intravascular contrast injections were performed in four animals and the performance characteristics of the device were evaluated in eight animals. In a comparative outcome study, 42 anesthetized and instrumented swine were randomized to receive LifeBelt or manual CPR. Ventricular fibrillation (VF) was induced electrically and was untreated for 7.5 min. After 7.5 min, countershocks were administered and chest compressions initiated. Pulseless electrical activity (PEA) was observed after one to three shocks in all animals. CPR was continued until restoration of spontaneous circulation (ROSC) or for 10 min after the first shock. If ROSC had not occurred within 5 min of beginning CPR, 0.01 mg/kg of epinephrine (adrenaline) was administered. During CPR, peak systolic aortic pressure (Ao), diastolic coronary perfusion pressure (CPP-diastolic aortic minus diastolic right pressure) and end-tidal CO(2) were measured.

RESULTS

Angiographic studies demonstrated cardiac compression as the mechanism of blood flow. Optimal performance, determined by coronary perfusion pressure, was observed at a sternal force of 100-130 lb (45-59 kg). In the comparative trial, significant differences in the measured CPP were observed between LifeBelt and manual CPR both at 1 min (15+/-8 mmHg versus 10+/-6 mmHg, p<0.05) and 5 min (17+/-4 mmHg versus 13+/-7 mmHg, p<0.02) of chest compression. A greater (p<0.05) ETCO(2), a marker of cardiac output and systemic perfusion, was observed with LifeBelt CPR (20+/-7 mmHg) than with manual CPR (15+/-5 mmHg) at 1 min. Peak Ao pressures were not different between methods. With the device, 86% of animals were resuscitated compared to 76% in the manual group.

CONCLUSIONS

Blood flow with the LifeBelt device is primarily the result of cardiac compression. At a sternal force of 100-130 lb (45-59 kg), the device produces greater CPP than well-performed manual CPR during resuscitation from prolonged VF.

به صفحه فیس بوک ما بپیوندید

کاملترین پایگاه داده گیاهان دارویی با پشتیبانی علمی

  • به 55 زبان کار می کند
  • درمان های گیاهی با پشتوانه علم
  • شناسایی گیاهان توسط تصویر
  • نقشه GPS تعاملی - گیاهان را در مکان نشان دهید (به زودی)
  • انتشارات علمی مربوط به جستجوی خود را بخوانید
  • گیاهان دارویی را با توجه به اثرات آنها جستجو کنید
  • علایق خود را سازماندهی کنید و با تحقیقات اخبار ، آزمایشات بالینی و حق ثبت اختراع در جریان باشید

علامت یا بیماری را تایپ کنید و در مورد گیاهانی که ممکن است به شما کمک کنند ، بخوانید ، یک گیاه تایپ کنید و بیماری ها و علائمی را که در برابر آن استفاده می شود ، ببینید.
* کلیه اطلاعات براساس تحقیقات علمی منتشر شده است

Google Play badgeApp Store badge