Respiratory Care 2012-Mar

Comparing performance of three oscillating positive expiratory pressure devices at similar amplitude and frequencies of oscillations on displacement of mucus inside trachea during cough.

فقط کاربران ثبت نام شده می توانند مقالات را ترجمه کنند
ورود به سیستم / ثبت نام
پیوند در کلیپ بورد ذخیره می شود
Anpalaki J Ragavan

کلید واژه ها

خلاصه

BACKGROUND

Performance of Flutter® (Axcan Scandipharm Inc, Birmingham, AL), Acapella® (Smiths Medicals Inc, Rockland, MA) and Quake® (Thayer Medical, Tucson, AZ) were compared at similar frequencies and amplitudes of oscillations at nine angles of the device in clearing simulated mucus inside a tracheal model (trachea) oriented at three angles with or without simulated constrictions in airway upstream of trachea.

METHODS

Displacement of 0.4mL of simulated mucus prepared with viscoelastic properties similar to healthy individuals (syrup-like) or patients with COPD (gel-like) using locust bean gum(LBG) solution (0.38g LBG in 100mL water) cross-linked with 3mL or 12mL borax solution (0.02 molar), respectively were measured inside trachea during coughs of 300ms at low cough velocity (15±0.5m/s) generated using a computer controlled solenoid valve. Oscillations were superimposed on cough by connecting the oscillator device to the outlet of the trachea.

RESULTS

Frequency and amplitude of oscillations generated by Quake and Acapella and resulting mucus displacement were independent of angle of oscillator, while amplitude of oscillations and resulting mucus displacement generated by Flutter, increased up to 30o upward and 20o downward angles of Flutter from horizontal but decreased significantly thereafter. Displacement with Quake increased significantly with frequencies of oscillations up to 25 Hz and decreased thereafter but increased with amplitudes of oscillations up to 22±4.7 m/s. Quake showed significantly larger displacements than Flutter and Acapella at equal frequencies and amplitudes (p<0.05). Displacements were significantly larger with trachea positioned 30o upwards than horizontal or 20o downwards (p<0.0001). Displacement was the greatest for gel-like mucus than syrup-like (p<0.0001). Airway constrictions upstream resulted in enhanced displacement of mucus (p<0.0001).

CONCLUSIONS

Mucus clearance can be significantly enhanced by coughing through oscillating positive expiratory devices that generate high amplitude oscillations at moderate frequencies, increasing frontal depths of mucus facing airflow and slightly increasing resistance to airflow in airways in COPD patients.

به صفحه فیس بوک ما بپیوندید

کاملترین پایگاه داده گیاهان دارویی با پشتیبانی علمی

  • به 55 زبان کار می کند
  • درمان های گیاهی با پشتوانه علم
  • شناسایی گیاهان توسط تصویر
  • نقشه GPS تعاملی - گیاهان را در مکان نشان دهید (به زودی)
  • انتشارات علمی مربوط به جستجوی خود را بخوانید
  • گیاهان دارویی را با توجه به اثرات آنها جستجو کنید
  • علایق خود را سازماندهی کنید و با تحقیقات اخبار ، آزمایشات بالینی و حق ثبت اختراع در جریان باشید

علامت یا بیماری را تایپ کنید و در مورد گیاهانی که ممکن است به شما کمک کنند ، بخوانید ، یک گیاه تایپ کنید و بیماری ها و علائمی را که در برابر آن استفاده می شود ، ببینید.
* کلیه اطلاعات براساس تحقیقات علمی منتشر شده است

Google Play badgeApp Store badge