Persian
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Journal of Cataract and Refractive Surgery 2006-Feb

Comparison of thermal features associated with 2 phacoemulsification machines.

فقط کاربران ثبت نام شده می توانند مقالات را ترجمه کنند
ورود به سیستم / ثبت نام
پیوند در کلیپ بورد ذخیره می شود
Jason P Brinton
Wesley Adams
Rajiv Kumar
Randall J Olson

کلید واژه ها

خلاصه

OBJECTIVE

To determine the thermal characteristics of the Legacy Advantec and Sovereign WhiteStar phacoemulsification machines during different clinically relevant scenarios.

METHODS

In vitro study.

METHODS

In water, temperature was recorded continuously on the sleeve in an artificial chamber, and the increase in temperature over baseline after 60 seconds of ultrasound was determined. This was done for continuous ultrasound, 50 ms on and 50 ms off (pulse), 6 ms on and 12 ms off (WhiteStar; Sovereign only) with aspiration blocked and not blocked, and with 100 g and 200 g weights suspended from the sleeve.

RESULTS

Comparing temperature increase per 20% machine power increments, Sovereign ran hotter than Legacy Advantec for continuous ultrasound (2.31x) and pulse (2.23x). Blocking aspiration increased temperature over the unblocked state. Pulsing decreased temperature by 51% (Legacy Advantec, pulse), 52% (Sovereign, pulse), and 64% (WhiteStar). Weights had much more effect on the Legacy Advantec: 3.5 times more going from baseline to 100 g weights and 3.2 times more going from 100 to 200 g weights. For all these comparisons, the P value was less than 0.0001.

CONCLUSIONS

The machines behaved fundamentally differently, with the Legacy Advantec controlling stroke length and Sovereign controlling a fixed power at any setting. Therefore, workload had a much bigger impact on Legacy Advantec thermal characteristics. Pulsing decreased heat produced directly related to the duty cycle. The most dangerous incision burn scenario is with continuous ultrasound, aspiration blocked, and a heavy workload.

به صفحه فیس بوک ما بپیوندید

کاملترین پایگاه داده گیاهان دارویی با پشتیبانی علمی

  • به 55 زبان کار می کند
  • درمان های گیاهی با پشتوانه علم
  • شناسایی گیاهان توسط تصویر
  • نقشه GPS تعاملی - گیاهان را در مکان نشان دهید (به زودی)
  • انتشارات علمی مربوط به جستجوی خود را بخوانید
  • گیاهان دارویی را با توجه به اثرات آنها جستجو کنید
  • علایق خود را سازماندهی کنید و با تحقیقات اخبار ، آزمایشات بالینی و حق ثبت اختراع در جریان باشید

علامت یا بیماری را تایپ کنید و در مورد گیاهانی که ممکن است به شما کمک کنند ، بخوانید ، یک گیاه تایپ کنید و بیماری ها و علائمی را که در برابر آن استفاده می شود ، ببینید.
* کلیه اطلاعات براساس تحقیقات علمی منتشر شده است

Google Play badgeApp Store badge