Diabetic ketoacidosis: difference between potassium determined by blood gas analysis versus plasma measurement.
کلید واژه ها
خلاصه
OBJECTIVE
To evaluate the accuracy of potassium concentrations measured by blood gas analysis (PBG) compared with laboratory serum potassium (LSP), in the initial care of patients with diabetic ketoacidosis (DKA).
METHODS
Fifty three patients with diabetes mellitus were evaluated in a retrospective analysis. PBG was carried out using the Radiometer ABL 700 (Radiometer Copenhagen(®)), and results were compared with LSP ADVIA 1650 Chemistry system (Siemens(®)), the gold standard method. Both methods are based on potentiometry.
RESULTS
Mean PBG was 3.66 mmol/L and mean LSP was 4.79 mmol/L. Mean difference between PBG and LSP was -1.13 mmol/L (p < 0.0005, 95% CI, -1.39 to -0,86). Lin concordance correlation coefficient was rc = 0.28 (95% CIb, 0.10 to 0.45), demonstrating low concordance between the methods.
CONCLUSIONS
Although PBG measurement is faster and easier, it should not be used as a surrogate for LSP in the clinical treatment of DKA.