Critical Care Medicine 2000-May

Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters.

فقط کاربران ثبت نام شده می توانند مقالات را ترجمه کنند
ورود به سیستم / ثبت نام
پیوند در کلیپ بورد ذخیره می شود
B Bohnhorst
C S Peter
C F Poets

کلید واژه ها

خلاصه

OBJECTIVE

Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact. Newly developed instruments have lower false alarm rates. We wanted to know whether this is achieved at the expense of an increased proportion of false negative alarms such as missed or delayed identification of hypoxemia and/or bradycardia.

METHODS

Observational study.

METHODS

Neonatal intensive care unit.

METHODS

A total of 17 unsedated preterm infants (median gestational age at birth, 25 wks; range, 24-30 wks).

METHODS

Long-term recordings of transcutaneous partial pressure of oxygen (P(Tc)O2), heart rate, pulse oximeter saturation (SpO2), and pulse rate from a conventional oximeter and two new generation oximeters.

METHODS

Recordings were analyzed for episodes with P(Tc)O2 <40 torr or with heart rate <80 beats/min for >5 secs. Hypoxemia was considered identified if SpO2 had fallen to <85% within 2 mins of P(Tc)O2 reaching 40 torr, and bradycardia was considered identified if pulse rate had fallen to <80 beats/min within 2 mins of the heart rate reaching this threshold.

RESULTS

A total of 202 falls in P(Tc)O2 to <40 torr occurred; 174 (86%) were identified by all three oximeters. Of the remaining episodes, manual analysis of red and infrared absorption signals confirmed that SpO2 had indeed been <85% for > or =10 secs in 11 episodes; therefore, these episodes should have been identified by all three oximeters. None of these had been missed by the conventional oximeter, but 10 (5.4% of the total) were missed by one of the new generation instruments (Nellcor), and one (0.5%) was missed by the other (Masimo). Of 54 bradycardias, only 14 were identified by all three oximeters; 17 (32%) were missed by the conventional, 37 (69%) by the Nellcor, and 4 (7%) by the Masimo instrument.

CONCLUSIONS

One of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias. It thus appears that this instrument's reduced false alarm rate is achieved at the expense of an unreliable and/or delayed identification of hypoxemia and bradycardia. The other instrument identified both conditions equally as or more reliably than a conventional pulse oximeter.

به صفحه فیس بوک ما بپیوندید

کاملترین پایگاه داده گیاهان دارویی با پشتیبانی علمی

  • به 55 زبان کار می کند
  • درمان های گیاهی با پشتوانه علم
  • شناسایی گیاهان توسط تصویر
  • نقشه GPS تعاملی - گیاهان را در مکان نشان دهید (به زودی)
  • انتشارات علمی مربوط به جستجوی خود را بخوانید
  • گیاهان دارویی را با توجه به اثرات آنها جستجو کنید
  • علایق خود را سازماندهی کنید و با تحقیقات اخبار ، آزمایشات بالینی و حق ثبت اختراع در جریان باشید

علامت یا بیماری را تایپ کنید و در مورد گیاهانی که ممکن است به شما کمک کنند ، بخوانید ، یک گیاه تایپ کنید و بیماری ها و علائمی را که در برابر آن استفاده می شود ، ببینید.
* کلیه اطلاعات براساس تحقیقات علمی منتشر شده است

Google Play badgeApp Store badge