Journal of Burn Care and Research

The reliability and concurrent validity of the figure-of-eight method of measuring hand edema in patients with burns.

فقط کاربران ثبت نام شده می توانند مقالات را ترجمه کنند
ورود به سیستم / ثبت نام
پیوند در کلیپ بورد ذخیره می شود
William S Dewey
Travis L Hedman
Ted T Chapman
Steven E Wolf
John B Holcomb

کلید واژه ها

خلاصه

OBJECTIVE

Water volumetry is considered the "gold standard" for hand edema assessment. This technique requires considerable time, staff, and specialized equipment. The figure-of-eight method for hand edema assessment has been tested only in the orthopedic population. The objective of this study was to test the reliability and concurrent validity of the figure-of-eight method of measuring hand edema in the burn patient population.

METHODS

We conducted a prospective blinded study with 20 burned patients (33 edematous hands) admitted from February to May 2005. Two testers performed three separate blinded measurements on each edematous hand, using the figure-of-eight technique. A third tester performed two measurements, using water volumetry. An independent investigator recorded all measurements. Intratester and intertester reliability were analyzed. Concurrent validity was examined and compared with water volumetry measurements.

RESULTS

Intraclass correlation coefficients (ICC) for the intratester reliability of the figure-of-eight method were 0.96 for tester 1 and 0.97 for tester 2. The ICC for intertester reliability of the figure-of-eight measurements was 0.94. The intratester ICC for volumetric measurements was 0.99. Correlation coefficient (Pearson's) for tester 1 was 0.83 (P < .01), and for tester 2, 0.89 (P < .01).

CONCLUSIONS

The figure-of-eight technique is a reliable and valid measurement tool for measuring hand edema. This technique is a more clinically feasible tool than water volumetry in the burn patient population.

به صفحه فیس بوک ما بپیوندید

کاملترین پایگاه داده گیاهان دارویی با پشتیبانی علمی

  • به 55 زبان کار می کند
  • درمان های گیاهی با پشتوانه علم
  • شناسایی گیاهان توسط تصویر
  • نقشه GPS تعاملی - گیاهان را در مکان نشان دهید (به زودی)
  • انتشارات علمی مربوط به جستجوی خود را بخوانید
  • گیاهان دارویی را با توجه به اثرات آنها جستجو کنید
  • علایق خود را سازماندهی کنید و با تحقیقات اخبار ، آزمایشات بالینی و حق ثبت اختراع در جریان باشید

علامت یا بیماری را تایپ کنید و در مورد گیاهانی که ممکن است به شما کمک کنند ، بخوانید ، یک گیاه تایپ کنید و بیماری ها و علائمی را که در برابر آن استفاده می شود ، ببینید.
* کلیه اطلاعات براساس تحقیقات علمی منتشر شده است

Google Play badgeApp Store badge