Neurosurgical Focus 2016-Sep

An improved reservoir for the flushing test to diagnose shunt insufficiency.

Vain rekisteröityneet käyttäjät voivat kääntää artikkeleita
Kirjaudu sisään Rekisteröidy
Linkki tallennetaan leikepöydälle
Hans-Georg Schlosser
Hans-Joachim Crawack
Christoph Miethke
Thoralf Knitter
Andreas Zeiner
Christian Sprung

Avainsanat

Abstrakti

OBJECTIVE Reservoirs integrated into hydrocephalus shunts are commonly used for the removal of CSF and for intra-ventricular pressure measurement. Pumping with the reservoir to diagnose shunt sufficiency is still a matter of controversy. The authors describe an improved flushing device and its characteristic features in vitro and in vivo. METHODS The flushing reservoir is constructed with a sapphire ball in a cage as a nonresistance valve to also enable the detection of distal occlusions. The most important reservoir parameters were investigated in vitro, simulating total and partial proximal and distal shunt occlusions. Then the expected advantages were assessed in vivo by evaluating the pump test data of 360 implanted reservoirs. The results were compared with those found in the literature. RESULTS The optimization of the technical parameters of the device, such as the high stroke volume in combination with moderate suction force, are obvious advantages compared with other flushing devices. Total occlusion of the ventricular catheter and the valve could be assessed with high certainty. The detection of a total obstruction of the peritoneal catheter or any partial obstruction is also possible, depending on its exact grade and location. CONCLUSIONS Shunt obstructions can be assessed using the pumping test. The reservoir construction presented here provides a clear enhancement of that diagnostic test.

Liity facebook-sivullemme

Täydellisin lääketieteellinen tietokanta tieteen tukemana

  • Toimii 55 kielellä
  • Yrttilääkkeet tieteen tukemana
  • Yrttien tunnistaminen kuvan perusteella
  • Interaktiivinen GPS-kartta - merkitse yrtit sijaintiin (tulossa pian)
  • Lue hakuusi liittyviä tieteellisiä julkaisuja
  • Hae lääkekasveja niiden vaikutusten perusteella
  • Järjestä kiinnostuksesi ja pysy ajan tasalla uutisista, kliinisistä tutkimuksista ja patenteista

Kirjoita oire tai sairaus ja lue yrtteistä, jotka saattavat auttaa, kirjoita yrtti ja näe taudit ja oireet, joita vastaan sitä käytetään.
* Kaikki tiedot perustuvat julkaistuun tieteelliseen tutkimukseen

Google Play badgeApp Store badge