Graefe's Archive for Clinical and Experimental Ophthalmology 2011-Jan

A disposable system for rapid purification of autologous plasmin as an adjunct to vitrectomy - performance and safety profile.

Seuls les utilisateurs enregistrés peuvent traduire des articles
Se connecter S'inscrire
Le lien est enregistré dans le presse-papiers
Martin Hermel
Wendelin Dailey
Michael Trese
Michael K Hartzer

Mots clés

Abstrait

BACKGROUND

The generation of an atraumatic posterior vitreous detachment (PVD), a common goal in vitreoretinal surgery, is a challenge particularly in children and young trauma patients. Plasmin has been proposed as a surgical adjunct to enzymatically generate a PVD. This study assesses the performance and safety of a new system for rapid purification of plasmin as an adjunct to vitrectomy.

METHODS

Plasminogen was isolated from human plasma by affinity chromatography using a disposable rapid purification kit, and activated to plasmin with streptokinase. Activities were assessed spectrophotometrically. For safety studies, 38 rabbits received intravitreal injections of one of the following compounds in 0.1 ml respectively: 4.7, 12.7 and 24 IU plasmin, 15 mg dextran, 4,100 U streptokinase, 500 μg ε-aminocaproic acid, 0.1 M potassium phosphate or balanced salt solution (BSS). Thirty min after injection, a two-port vitrectomy was performed. Rabbits were followed clinically and with bright flash electroretinography (ERG) for up to 9 months. The eyes were investigated by light and transmission electron microscopy.

RESULTS

The specific plasmin activity obtained from blood of healthy volunteers averaged 42.3 ± 6.6 IU/ml (range 21.6 IU/ml to 54.5 IU/ml). The identity and purity of the enzyme was confirmed by several methods. Clinically, a mild to moderate inflammatory response was seen in most eyes on day 1, but had disappeared by day 7. ERG showed moderate depressions of a- and b-wave amplitudes on day 2, particularly in the potassium phosphate (a: -29.16 ± 4.56, b: -21.23 ± 6.31), 4.7 (a: -34.38 ± 6.64, b: -26.66 ± 6.06) and 24 IU (a: -38.25 ± 4.05, b: -23.38 ± 4.29) plasmin groups, but also in the BSS- (a: -11.19 ± 21.78, b: -11.41 ± 15.47) and dextran- (a: -17.86 ± 14.18, b: -6.67 ± 18.14) treated eyes. ERG changes recovered during follow-up. One rabbit each from the 12.7 and the 24 IU plasmin groups showed a minimal discoloration of one medullary ray after 9 months. Histology did not reveal morphologic signs of toxicity.

CONCLUSIONS

The isolation system generated plasmin with a high degree of purity. A failure-mode analysis did not reveal significant risks of toxicity. A single preparation can provide a maximum dose of 10.9 IU/200 μl, the likely target clinical dose being 1.88 IU. Plasmin doses of at least 12.7 IU appear be safe when injected into rabbit eyes, followed by vitrectomy.

Rejoignez notre
page facebook

La base de données d'herbes médicinales la plus complète soutenue par la science

  • Fonctionne en 55 langues
  • Cures à base de plantes soutenues par la science
  • Reconnaissance des herbes par image
  • Carte GPS interactive - étiquetez les herbes sur place (à venir)
  • Lisez les publications scientifiques liées à votre recherche
  • Rechercher les herbes médicinales par leurs effets
  • Organisez vos intérêts et restez à jour avec les nouvelles recherches, essais cliniques et brevets

Tapez un symptôme ou une maladie et lisez des informations sur les herbes qui pourraient aider, tapez une herbe et voyez les maladies et symptômes contre lesquels elle est utilisée.
* Toutes les informations sont basées sur des recherches scientifiques publiées

Google Play badgeApp Store badge