Minimally invasive neurosurgery : MIN 2008-Dec

A laser-induced liquid jet catheter system: a novel endovascular device for rapid and reliable fibrinolysis in acute cerebral embolism.

Hanya pengguna terdaftar yang dapat menerjemahkan artikel
Masuk daftar
Tautan disimpan ke clipboard
T Hirano
A Nakagawa
T Ohki
H Uenohara
K Takayama
T Tominaga

Kata kunci

Abstrak

OBJECTIVE

Mechanical removal of intravascular clots in addition to administration of tissue plasminogen activator are both desirable for improved outcome in acute embolic stroke. We have developed a novel endovascular catheter system for rapid and reliable mechanical recanalization of cerebral embolisms with little or no requirement for fibrinolytic agents. Here, we describe the evaluation of this device in vitro.

METHODS

Pulsed liquid jets were generated and ejected from the catheter exit by accelerating cold physiological saline (4 degrees C, 40 mL/h) using the energy of a pulsed holmium:yttrium-aluminum-garnet (YAG) laser (3 Hz, 1.2 W). Accessibility beyond the tortuous cavernous portion of the internal carotid artery to the M1 and A1 regions was confirmed using a transparent model of the human cerebral artery. Mechanical characteristics of the liquid jets were evaluated with a high-speed camera. Liquid jets of physiological saline or urokinase solution (1,200 IU/mL) were exposed to artificial thrombi made of human blood under temperature monitoring. Remnants of thrombi were collected and incubated at 37 degrees C for 10 min for estimation of fibrinolysis rates.

RESULTS

The jet velocity (maximum: 5 m/s) was controlled by changing the laser energy. The fibrinolysis rates (mean+/-SD) after exposure to jets of saline or urokinase solution for 45 s were 62.2+/-16.4 and 94.0+/-3.4%, respectively, and were significantly better than the rate of 8.1+/-2.0% with administration of urokinase alone. The local temperature rise was less than 8 degrees C.

CONCLUSIONS

The results show that the laser-induced liquid jet catheter system may be a powerful tool for mechanical destruction of emboli and augmentation of the effect of fibrinolytic agents beyond the tortuous part of the internal carotid artery.

Bergabunglah dengan
halaman facebook kami

Database tanaman obat terlengkap yang didukung oleh sains

  • Bekerja dalam 55 bahasa
  • Pengobatan herbal didukung oleh sains
  • Pengenalan herbal melalui gambar
  • Peta GPS interaktif - beri tag herba di lokasi (segera hadir)
  • Baca publikasi ilmiah yang terkait dengan pencarian Anda
  • Cari tanaman obat berdasarkan efeknya
  • Atur minat Anda dan ikuti perkembangan berita, uji klinis, dan paten

Ketikkan gejala atau penyakit dan baca tentang jamu yang mungkin membantu, ketik jamu dan lihat penyakit dan gejala yang digunakan untuk melawannya.
* Semua informasi didasarkan pada penelitian ilmiah yang dipublikasikan

Google Play badgeApp Store badge