Indonesian
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Anesthesia and Analgesia 2011-Jun

Activated charcoal effectively removes inhaled anesthetics from modern anesthesia machines.

Hanya pengguna terdaftar yang dapat menerjemahkan artikel
Masuk daftar
Tautan disimpan ke clipboard
Nathaniel Birgenheier
Robert Stoker
Dwayne Westenskow
Joseph Orr

Kata kunci

Abstrak

BACKGROUND

If a malignant hyperthermia-susceptible patient is to receive an anesthetic, an anesthesia machine that has been used previously to deliver volatile anesthetics should be flushed with a high fresh gas flow. Conflicting results from previous studies recommend flush times that vary from 10 to 104 minutes. In a previously proposed alternative decontamination technique, other investigators placed an activated charcoal filter in the inspired limb of the breathing circuit.

METHODS

We placed activated charcoal filters on both the inspired and expired limbs of several contaminated anesthesia machines and measured the time needed to flush the machine so that the delivered concentrations of isoflurane, sevoflurane, and desflurane would be <5 parts per million (ppm). We next simulated the case for which malignant hyperthermia is diagnosed 90 minutes after induction of anesthesia and measured how well activated charcoal filters limit further exposure.

RESULTS

Activated charcoal filters decrease the concentration of volatile anesthetic delivered by a contaminated machine to an acceptable level in <2 minutes. The concentrations remained well below 5 ppm for at least 60 minutes. When malignant hyperthermia is diagnosed after induction of anesthesia, we found that with charcoal filters in place, the current anesthesia machine may be used for at least 67 minutes before the inspired concentration exceeds 5 ppm.

CONCLUSIONS

Activated charcoal filters provide an alternative approach to the 10 to 104 minutes of flushing that are normally required to prepare a machine that has been used previously to deliver a volatile anesthetic.

Bergabunglah dengan
halaman facebook kami

Database tanaman obat terlengkap yang didukung oleh sains

  • Bekerja dalam 55 bahasa
  • Pengobatan herbal didukung oleh sains
  • Pengenalan herbal melalui gambar
  • Peta GPS interaktif - beri tag herba di lokasi (segera hadir)
  • Baca publikasi ilmiah yang terkait dengan pencarian Anda
  • Cari tanaman obat berdasarkan efeknya
  • Atur minat Anda dan ikuti perkembangan berita, uji klinis, dan paten

Ketikkan gejala atau penyakit dan baca tentang jamu yang mungkin membantu, ketik jamu dan lihat penyakit dan gejala yang digunakan untuk melawannya.
* Semua informasi didasarkan pada penelitian ilmiah yang dipublikasikan

Google Play badgeApp Store badge