Journal of Heart Valve Disease 2010-Jul

In-vivo assessment of a novel polymer (SIBS) trileaflet heart valve.

Hanya pengguna terdaftar yang dapat menerjemahkan artikel
Masuk daftar
Tautan disimpan ke clipboard
Qiang Wang
Anthony J McGoron
Richard Bianco
Yasushi Kato
Leonard Pinchuk
Richard T Schoephoerster

Kata kunci

Abstrak

OBJECTIVE

A novel trileaflet polymer valve, which is a composite design of a biostable and biocompatible polymer poly(styrene-block-isobutylene-block-styrene) (SIBS) with an embedded reinforcement polyethylene terephthalate (PET) fabric, is being developed with the intention of providing a valve that has low thrombogenicity, high durability and favorable hemodynamic performance. The study aim was to investigate the biocompatibility and performance of this SIBS valve prototype under physiological loading conditions similar to humans, using a large-animal model.

METHODS

Four SIBS valves (two with surface modification using dimyristoyl phosphatidylcholine, DMPC), and two commercial Magna tissue valves, were implanted into sheep. Hemodynamic and blood chemistry measurements were performed periodically during the postoperative period. The explanted SIBS valves were extensively evaluated using macroscopic, histological, radiographical and scanning electron microscopy/energy-dispersive spectroscopy analysis.

RESULTS

Three animals, one with the DMPC-coated SIBS valve, and two with the Magna valves, reached the end of the study in satisfactory clinical condition, and were euthanized after 20 weeks. The other three animals (two with SIBS valves, one with a DMPC-coated SIBS valve) died at 6, 6.5, and 10 weeks due either to material failure or myocardial infarction. The explanted valves exhibited stent deformation and cracks on the leaflets, which exposed the underlying PET fabric and resulted in severe blood and tissue reactions. Extrinsic calcification was identified on the leaflets, and was associated with the regions of surface cracks.

CONCLUSIONS

The SIBS valve failed in animal testing because of material failure and calcification. The physical properties of SIBS must be improved in order to provide the structural integrity required for long-term in-vivo use in the form of a heart valve.

Bergabunglah dengan
halaman facebook kami

Database tanaman obat terlengkap yang didukung oleh sains

  • Bekerja dalam 55 bahasa
  • Pengobatan herbal didukung oleh sains
  • Pengenalan herbal melalui gambar
  • Peta GPS interaktif - beri tag herba di lokasi (segera hadir)
  • Baca publikasi ilmiah yang terkait dengan pencarian Anda
  • Cari tanaman obat berdasarkan efeknya
  • Atur minat Anda dan ikuti perkembangan berita, uji klinis, dan paten

Ketikkan gejala atau penyakit dan baca tentang jamu yang mungkin membantu, ketik jamu dan lihat penyakit dan gejala yang digunakan untuk melawannya.
* Semua informasi didasarkan pada penelitian ilmiah yang dipublikasikan

Google Play badgeApp Store badge