Journal of Aerosol Medicine and Pulmonary Drug Delivery 2017-Apr

A Novel In-Line Delivery System to Administer Dry Powder Mannitol to Mechanically Ventilated Patients.

Solo gli utenti registrati possono tradurre articoli
Entra registrati
Il collegamento viene salvato negli appunti
Benny Feng
Patricia Tang
Sharon Shui Yee Leung
Jayesh Dhanani
Hak-Kim Chan

Parole chiave

Astratto

BACKGROUND

Mechanically ventilated patients commonly suffer from ventilator-associated pneumonia, hypoxemia, and other lower respiratory tract infection as a result of pathogen colonization and poor sputum clearance. Consequently, there is a high rate of morbidity and mortality in these patients. Dry powder mannitol increases sputum clearance, and therefore, we developed a system to administer it to mechanically ventilated patients without disconnection from the ventilator.

METHODS

The inspiratory line from a ventilator was split by using a three-way valve into two parallel lines where one contains a humidifier for normal breathing cycle and the other line contains a dry powder inhaler (Osmohaler™). The inspiratory air went through the dry powder line and aerosolized the mannitol powder only when its administration to a patient is required. We determined the delivered dose and particle size distributions of emitted aerosols in vitro from 9.5 mm endotracheal and 7.5 mm tracheostomy tubes, with inspiratory airflow of 60, 70, and 80 L/min.

RESULTS

This novel setup was able to deliver 24.6% ± 3.33% of the 160 mg loaded dose mannitol powder (4 × 40 mg capsules) and 26.7% ± 2.19% of the 320 mg dose (4 × 80 mg capsules) when the endotracheal tube was used. With the shorter tracheostomy tube, the delivery dose increased to 35.6% ± 3.01% and 39.5% ± 2.04% of the 160 and 320 mg doses, respectively. The volume median diameters of the aerosols were in the respirable range with the largest value being 5.17 ± 0.87 μm.

CONCLUSIONS

This delivery system has been shown to consistently deliver a high respirable dose of mannitol powder. Since this setup does not require disconnection of patients from the ventilator, it is safer for hypoxemic patients and easier to be adapted in a real clinical use.

Unisciti alla nostra
pagina facebook

Il database di erbe medicinali più completo supportato dalla scienza

  • Funziona in 55 lingue
  • Cure a base di erbe sostenute dalla scienza
  • Riconoscimento delle erbe per immagine
  • Mappa GPS interattiva - tagga le erbe sul luogo (disponibile a breve)
  • Leggi le pubblicazioni scientifiche relative alla tua ricerca
  • Cerca le erbe medicinali in base ai loro effetti
  • Organizza i tuoi interessi e tieniti aggiornato sulle notizie di ricerca, sperimentazioni cliniche e brevetti

Digita un sintomo o una malattia e leggi le erbe che potrebbero aiutare, digita un'erba e osserva le malattie ei sintomi contro cui è usata.
* Tutte le informazioni si basano su ricerche scientifiche pubblicate

Google Play badgeApp Store badge