Biomedical Signal Processing and Control 2014-Sep

Statistical evaluation of reproducibility of automated ECG measurements: an example from arrhythmogenic right ventricular dysplasia/cardiomyopathy clinic.

登録ユーザーのみが記事を翻訳できます
ログインサインアップ
リンクがクリップボードに保存されます
Timothy Huang
Cynthia A James
Crystal Tichnell
Brittney Murray
Joel Xue
Hugh Calkins
Larisa G Tereshchenko

キーワード

概要

BACKGROUND

Arrhythmogenic right ventricular dysplasia/cardiomyopathy (ARVD/C) is characterized by delay in depolarization of the right ventricle, detected by prolonged terminal activation duration (TAD) in V1-V3. However, manual ECG measurements have shown moderate-to-low intra- and inter-reader agreement. The goal of this study was to assess reproducibility of automated ECG measurements in the right precordial leads.

METHODS

Pairs of ECGs recorded in the same day from Johns Hopkins ARVD/C Registry participants [n=247, mean age 35.2±15.6 y, 58% men, 92% whites, 11(4.5%) with definite ARVD/C] were retrospectively analyzed. QRS duration, intrinsicoid deflection, TAD, and T-wave amplitude in the right precordial leads, as well as averaged across all leads QRS duration, QRS axis, T axis, QTc interval, and heart rate was measured automatically, using 12SL TM algorithm (GE Healthcare, Wauwatosa, WI, USA). Intrinsicoid deflection was measured as the time from QRS complex onset to the alignment point of the QRS complex. TAD was calculated as the difference between QRS duration and intrinsicoid in V1, V2, V3. Reproducibility was quantified by Bland-Altman analysis (bias with 95% limits of agreement), Lin's concordance coefficient, and Bradley-Blackwood procedure.

RESULTS

Bland-Altman analysis revealed satisfactory reproducibility of tested parameters. V1 QRS duration bias was -0.10ms [95% limits of agreement -12.77 to 12.56ms], V2 QRS duration bias -0.09ms [-11.13 to 10.96ms]; V1 TAD bias 0.14ms [-13.23 to 13.51ms], V2 TAD bias 0.008ms [-12.42 to 12.44ms].

CONCLUSIONS

Comprehensive statistical evaluation of reproducibility of automated ECG measurements is important for appropriate interpretation of ECG. Automated ECG measurements are reproducible to within 25%.

Facebookページに参加する

科学に裏打ちされた最も完全な薬草データベース

  • 55の言語で動作します
  • 科学に裏打ちされたハーブ療法
  • 画像によるハーブの認識
  • インタラクティブGPSマップ-場所にハーブをタグ付け(近日公開)
  • 検索に関連する科学出版物を読む
  • それらの効果によって薬草を検索する
  • あなたの興味を整理し、ニュース研究、臨床試験、特許について最新情報を入手してください

症状や病気を入力し、役立つ可能性のあるハーブについて読み、ハーブを入力して、それが使用されている病気や症状を確認します。
*すべての情報は公開された科学的研究に基づいています

Google Play badgeApp Store badge