Current Eye Research 2014-Sep

Tonometry in corneal edema after cataract surgery: rebound versus goldmann applanation tonometry.

登録ユーザーのみが記事を翻訳できます
ログインサインアップ
リンクがクリップボードに保存されます
Matthias Fuest
Nikolaos Mamas
Peter Walter
Niklas Plange

キーワード

概要

OBJECTIVE

Goldmann applanation tonometry (GAT) is known to be influenced by corneal properties, whereas the effect of corneal edema on intraocular pressure (IOP) measurement using Rebound tonometry (RT) is not clear. In this study, IOP was measured before and after cataract surgery using GAT and RT to investigate differences between methods and the effect of postsurgical corneal edema.

METHODS

Thirty patients with cataract were included in a prospective study. IOP was measured using GAT and RT (Icare PRO, Tiolat Oy, Finland) before and after one day of phacoemulsification. Central corneal thickness (CCT) was determined before and after surgery (Pentacam, Oculus, Germany).

RESULTS

CCT increased significantly after surgery by 52.2 ± 35.1 µm (p < 0.0001). IOP values of GAT and RT were significantly correlated before surgery (r = 0.65, p < 0.0001) and after surgery (r = 0.94, p < 0.0001). No significant correlation of GAT and RT to CCT before surgery (GAT: r = 0.18, p = 0.36; RT: r = 0.03, p = 0.87) or after surgery (GAT: r = -0.08, p = 0.69; RT: r = 0.17, p = 0.37) was found. The mean difference between GAT and RT (GAT-RT) was -1.0 ± 2.9 mmHg (range -6.2 to 6.4 mmHg, p = 0.08) before surgery and -1.1 ± 2.2 mmHg (range -5.1 to 4.8 mmHg, p = 0.01) after surgery. Before surgery, a positive correlation of GAT-RT with mean IOP of both methods is apparent (r = 0.45, p = 0.01). After surgery, no such correlation was found (r = -0.08, p = 0.69).

CONCLUSIONS

RT underestimates IOP in higher values and overestimates IOP in lower values compared to GAT before surgery. This is not apparent in eyes with corneal edema. Large differences between methods may be present in individual patients with and without corneal edema.

Facebookページに参加する

科学に裏打ちされた最も完全な薬草データベース

  • 55の言語で動作します
  • 科学に裏打ちされたハーブ療法
  • 画像によるハーブの認識
  • インタラクティブGPSマップ-場所にハーブをタグ付け(近日公開)
  • 検索に関連する科学出版物を読む
  • それらの効果によって薬草を検索する
  • あなたの興味を整理し、ニュース研究、臨床試験、特許について最新情報を入手してください

症状や病気を入力し、役立つ可能性のあるハーブについて読み、ハーブを入力して、それが使用されている病気や症状を確認します。
*すべての情報は公開された科学的研究に基づいています

Google Play badgeApp Store badge