Korean
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Ophthalmologe 2010-Sep

[Clinical results of the aspheric intraocular lens FY-60AD (Hoya) with particular respect to decentration and tilt].

등록 된 사용자 만 기사를 번역 할 수 있습니다.
로그인 / 가입
링크가 클립 보드에 저장됩니다.
U Mester
S Heinen
H Kaymak

키워드

요약

BACKGROUND

Aspheric intraocular lenses (IOLs) aim to improve visual function and particularly contrast vision by neutralizing spherical aberration. One drawback of such IOLs is the enhanced sensitivity to decentration and tilt, which can deteriorate image quality.

METHODS

A total of 30 patients who received bilateral phacoemulsification before implantation of the aspheric lens FY-60AD (Hoya) were included in a prospective study. In 25 of the patients (50 eyes) the following parameters could be assessed 3 months after surgery: visual acuity, refraction, contrast sensitivity, pupil size, wavefront errors and decentration and tilt using a newly developed device.

RESULTS

The functional results were very satisfying and comparable to results gained with other aspheric IOLs. The mean refraction was sph + 0.1 D (±0.7 D) and cyl 0.6 D (±0.8 D). The spherical equivalent was −0.2 D (±0.6 D). Wavefront measurements revealed a good compensation of the corneal spherical aberration but vertical and horizontal coma also showed opposing values in the cornea and IOL. The assessment of the lens position using the Purkinje meter demonstrated uncritical amounts of decentration and tilt. The mean amount of decentration was 0.2 mm±0.2 mm in the horizontal and vertical directions. The mean amount of tilt was 4.0±2.1° in horizontal and 3.0±2.5° in vertical directions.

CONCLUSIONS

In a normal dioptric power range the aspheric IOL FY-60AD compensates the corneal spherical aberration very well with only minimal decentration. The slight tilt is symmetrical in both eyes and corresponds to the position of the crystalline lens in young eyes. This may contribute to our findings of compensated corneal coma.

페이스 북
페이지에 가입하세요

과학이 뒷받침하는 가장 완벽한 약초 데이터베이스

  • 55 개 언어로 작동
  • 과학이 뒷받침하는 약초 치료제
  • 이미지로 허브 인식
  • 인터랙티브 GPS지도-위치에 허브 태그 지정 (출시 예정)
  • 검색과 관련된 과학 출판물 읽기
  • 효과로 약초 검색
  • 관심사를 정리하고 뉴스 연구, 임상 실험 및 특허를 통해 최신 정보를 확인하세요.

증상이나 질병을 입력하고 도움이 될 수있는 약초에 대해 읽고 약초를 입력하고 사용되는 질병과 증상을 확인합니다.
* 모든 정보는 발표 된 과학 연구를 기반으로합니다.

Google Play badgeApp Store badge