Journal of Cataract and Refractive Surgery 2004-May

Legacy AdvanTec and Sovereign WhiteStar: a wound temperature study.

등록 된 사용자 만 기사를 번역 할 수 있습니다.
로그인 / 가입
링크가 클립 보드에 저장됩니다.
Randall J Olson
Yian Jin
Gus Kefalopoulos
Jason Brinton

키워드

요약

OBJECTIVE

To assess the wound temperature of the Sovereign WhiteStar (S-WS) (AMO) and Legacy AdvanTec (L-ADV) (Alcon) phacoemulsification systems.

METHODS

John A. Moran Eye Center, Health Sciences Center, University of Utah, Salt Lake City, Utah, USA.

METHODS

Phacoemulsification using 20-gauge, 30-degree straight tips with the L-ADV and S-WS systems was performed in fresh cadaver eyes. The power was set at 50%, and aspiration was 12 mL/min; the L-ADV was run at 15 pulses per second (pps) and the S-WS at WS CF (6 milliseconds on, 12 milliseconds off). Temperature was measured at 5-second intervals for 60 seconds using a microthermistor placed in the wound. The phaco tip was angled 30 degrees to increase wound-tissue contact. At 10 seconds, the flow was clamped to simulate occlusion. Five runs, sleeved and unsleeved, were averaged for the phaco tip.

RESULTS

The mean temperature was significantly higher with L-ADV than with S-WS (from 10 seconds on in the sleeved condition and from 5 seconds on in the unsleeved condition after the aspiration line was clamped). In 2 of 5 sleeved runs and 4 of 5 unsleeved runs, the L-ADV handpiece decreased power as the temperature increased. One run with the L-ADV with sleeve showed signs of wound burn. There were no signs of wound burn with the S-WS. The highest temperature recorded was 57.5 degrees C with the L-ADV and 38.6 degrees C with the S-WS. Power tests showed L-ADV protected the stroke length and S-WS had constant power except in air where it increased power.

CONCLUSIONS

There was less increase in wound temperature over time with the S-WS than with the L-ADV system in sleeved and unsleeved simulated surgery in human eye-bank eyes. Because the ultrasound handpieces respond differently under different load scenarios, meaningful comparisons of pulsing features are difficult to create.

페이스 북
페이지에 가입하세요

과학이 뒷받침하는 가장 완벽한 약초 데이터베이스

  • 55 개 언어로 작동
  • 과학이 뒷받침하는 약초 치료제
  • 이미지로 허브 인식
  • 인터랙티브 GPS지도-위치에 허브 태그 지정 (출시 예정)
  • 검색과 관련된 과학 출판물 읽기
  • 효과로 약초 검색
  • 관심사를 정리하고 뉴스 연구, 임상 실험 및 특허를 통해 최신 정보를 확인하세요.

증상이나 질병을 입력하고 도움이 될 수있는 약초에 대해 읽고 약초를 입력하고 사용되는 질병과 증상을 확인합니다.
* 모든 정보는 발표 된 과학 연구를 기반으로합니다.

Google Play badgeApp Store badge