Latvian
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Infection Control and Hospital Epidemiology 2004-Mar

Epidemic parenteral exposure to volatile sulfur-containing compounds at a hemodialysis center.

Rakstu tulkošanu var veikt tikai reģistrēti lietotāji
Ielogoties Reģistrēties
Saite tiek saglabāta starpliktuvē
Dejana Selenic
Francisco Alvarado-Ramy
Mathew Arduino
Stacey Holt
Fred Cardinali
Benjamin Blount
Jeff Jarrett
Forrest Smith
Neil Altman
Charlotte Stahl

Atslēgvārdi

Abstrakts

OBJECTIVE

To determine the cause of acute illness on August 30, 2000, among patients at an outpatient dialysis center (center A).

METHODS

We performed a cohort study of all patients receiving dialysis on August 30, 2000; reviewed dialysis procedures; and analyzed dialysis water samples using microbiologic and chemical assays.

METHODS

Dialysis center (center A).

METHODS

A case-patient was defined as a patient who developed chills within 5 hours after starting hemodialysis at center A on August 30, 2000.

RESULTS

Sixteen (36%) of 44 patients at center A met the case definition. All case-patients were hospitalized; 2 died. Besides chills, 15 (94%) of the case-patients experienced nausea; 12 (75%), vomiting; and 4 (25%), fever. Illness was more frequent on the second than the first dialysis shift (16 of 20 vs 0 of 24, P < .001); no other risk factors were identified. The center's water treatment system had received inadequate maintenance and disinfection and a sulfurous odor was noted during sampling of the water from the reverse osmosis (RO) unit. The water had elevated bacterial counts. Volatile sulfur-containing compounds (ie, methanethiol, carbon disulfide, dimethyldisulfide, and sulfur dioxide) were detected by gas chromatography and mass spectrometry in 8 of 12 water samples from the RO unit and in 0 of 28 samples from other areas (P < .001). Results of tests for heavy metals and chloramines were within normal limits.

CONCLUSIONS

Parenteral exposure to volatile sulfur-containing compounds, produced under anaerobic conditions in the RO unit, could have caused the outbreak. This investigation demonstrates the importance of appropriate disinfection and maintenance of water treatment systems in hemodialysis centers.

Pievienojieties mūsu
facebook lapai

Vispilnīgākā ārstniecības augu datu bāze, kuru atbalsta zinātne

  • Darbojas 55 valodās
  • Zāļu ārstniecības līdzekļi, kurus atbalsta zinātne
  • Garšaugu atpazīšana pēc attēla
  • Interaktīva GPS karte - atzīmējiet garšaugus atrašanās vietā (drīzumā)
  • Lasiet zinātniskās publikācijas, kas saistītas ar jūsu meklēšanu
  • Meklēt ārstniecības augus pēc to iedarbības
  • Organizējiet savas intereses un sekojiet līdzi jaunumiem, klīniskajiem izmēģinājumiem un patentiem

Ierakstiet simptomu vai slimību un izlasiet par garšaugiem, kas varētu palīdzēt, ierakstiet zāli un redziet slimības un simptomus, pret kuriem tā tiek lietota.
* Visa informācija ir balstīta uz publicētiem zinātniskiem pētījumiem

Google Play badgeApp Store badge