Vox Sanguinis 2001-Jan

Automated blood-mixing devices still fail to mix at low bleeding rates.

Зөвхөн бүртгэлтэй хэрэглэгчид л нийтлэл орчуулах боломжтой
Нэвтрэх / Бүртгүүлэх
Холбоосыг санах ойд хадгалдаг
D de Korte
H A Veldman

Түлхүүр үгс

Хураангуй

OBJECTIVE

The study evaluated mixing performance for various commercial blood-mixing devices at different flow rates.

METHODS

A glycerol solution with a density equal to blood was pumped into a blood bag (mounted on a mixer) that contained 70 ml of Toluidin Blue in citrate-phosphate-dextrose (CPD). After 500 ml of glycerol solution had been pumped into the blood bag, the bag was emptied in fractions and the absorbance at 640 nm of each fraction (expressed as percentage of mixture absorbance) was plotted against the fraction number.

RESULTS

At a flow rate of 30 ml/min, the curves were very sigmoid, with high initial values that decreased during collection of the fractions. The absorbance values of the fractions collected later (after approximately 250-300 ml) were 10-50% (i.e. percentage of absorbance for complete mixing) for the different mixers, indicating that the CPD solution was incompletely mixed with the incoming solution, resulting in a lower-than-expected 'anticoagulant' content for the fractions collected later. At a flow rate of 50 ml/min the mixing was improved, but only at 75 ml/min did all mixers show relatively good mixing. Manual mixing, by kneading the bag three times/min, gave at all rates an almost ideal mixing curve, 105% initially to 90% in the final fractions.

CONCLUSIONS

At relatively high bleeding rates, all mixers performed well, although complete mixing was only obtained with manual kneading. However, most blood-mixing devices still fail to mix efficiently at normal and low bleeding rates. Although the minimal degree of necessary mixing is unknown, further optimization of mixing devices seems warranted.

Манай facebook
хуудсанд нэгдээрэй

Шинжлэх ухаанаар баталгаажсан эмийн өвс ургамлын бүрэн мэдээллийн сан

  • 55 хэл дээр ажилладаг
  • Шинжлэх ухааны үндэслэсэн ургамлын гаралтай эдгэрэлт
  • Ургамлыг дүрсээр таних
  • Интерактив GPS газрын зураг - эмийн ургамлыг байршлаар нь тэмдэглэнэ (удахгүй)
  • Хайлттай холбоотой шинжлэх ухааны нийтлэлүүдийг уншина уу
  • Эмийн өвсийг үр нөлөөгөөр нь хайж олох
  • Мэдээллийн судалгаа, клиник туршилт, патентыг цаг тухайд нь сонирхож, зохион байгуул

Шинж тэмдэг эсвэл өвчний талаар бичиж, тус болох ургамлын талаар уншиж, өвслөг ургамлыг бичиж, өвчний эсрэг шинж тэмдгийг үзээрэй.
* Бүх мэдээлэл нь хэвлэгдсэн эрдэм шинжилгээний судалгаанд үндэслэсэн болно

Google Play badgeApp Store badge