Vascular Cell 2014

Evaluation of side effects of radiofrequency capacitive hyperthermia with magnetite on the blood vessel walls of tumor metastatic lesion surrounding the abdominal large vessels: an agar phantom study.

Зөвхөн бүртгэлтэй хэрэглэгчид л нийтлэл орчуулах боломжтой
Нэвтрэх / Бүртгүүлэх
Холбоосыг санах ойд хадгалдаг
Noriyasu Kawai
Daichi Kobayashi
Takahiro Yasui
Yukihiro Umemoto
Kentaro Mizuno
Atsushi Okada
Keiichi Tozawa
Takeshi Kobayashi
Kenjiro Kohri

Түлхүүр үгс

Хураангуй

BACKGROUND

Magnetite used in an 8-MHz radiofrequency (RF) capacitive heating device can increase the temperature of a specific site up to 45°C. When treating a metastatic lesion around large abdominal vessels via hyperthermia with magnetite, heating-induced adverse effects on these vessels need to be considered. Therefore, this study examined hyperthermia-induced damage to blood vessel walls in vitro.

METHODS

A large agar phantom with a circulatory system consisting of a swine artery and vein connected to a peristaltic pump was prepared. The blood vessels were placed on the magnetite-containing agar piece. Heating was continued for 30 min at 45°C. After heating, a histological study for injury to the blood vessels was performed.

RESULTS

The inner membrane temperature did not reach 45°C due to the cooling effect of the blood flow. In the heated vessels, vascular wall collagen degenerated and smooth muscle cells were narrowed; however, no serious changes were noted in the vascular endothelial cells or vascular wall elastic fibers. The heated vessel wall was not severely damaged; this was attributed to cooling by the blood flow.

CONCLUSIONS

Our findings indicate that RF capacitive heating therapy with magnetite may be used for metastatic lesions without injuring the surrounding large abdominal vessels.

Манай facebook
хуудсанд нэгдээрэй

Шинжлэх ухаанаар баталгаажсан эмийн өвс ургамлын бүрэн мэдээллийн сан

  • 55 хэл дээр ажилладаг
  • Шинжлэх ухааны үндэслэсэн ургамлын гаралтай эдгэрэлт
  • Ургамлыг дүрсээр таних
  • Интерактив GPS газрын зураг - эмийн ургамлыг байршлаар нь тэмдэглэнэ (удахгүй)
  • Хайлттай холбоотой шинжлэх ухааны нийтлэлүүдийг уншина уу
  • Эмийн өвсийг үр нөлөөгөөр нь хайж олох
  • Мэдээллийн судалгаа, клиник туршилт, патентыг цаг тухайд нь сонирхож, зохион байгуул

Шинж тэмдэг эсвэл өвчний талаар бичиж, тус болох ургамлын талаар уншиж, өвслөг ургамлыг бичиж, өвчний эсрэг шинж тэмдгийг үзээрэй.
* Бүх мэдээлэл нь хэвлэгдсэн эрдэм шинжилгээний судалгаанд үндэслэсэн болно

Google Play badgeApp Store badge