Pregnancy Hypertension 2018-Jan

Analytical validation of soluble fms-like tyrosine and placental growth factor assays on B·R·A·H·M·S KRYPTOR Compact Plus automated immunoassay platform.

Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
Siaw Li Chan
Sarosh Rana
Sireesha Chinthala
Saira Salahuddin
Kiang-Teck J Yeo

Sleutelwoorden

Abstract

BACKGROUND

Preeclampsia is one of the leading hypertensive disorders of pregnancy. Angiogenic biomarkers such as anti-angiogenic factor soluble fms-like tyrosine kinase 1 (sFlt1) and pro-angiogenic factor placental growth factor (PlGF) are involved in the pathophysiology of preeclampsia.

OBJECTIVE

The aim of this study is to validate the analytical performance of sFlt1 and PlGF on the B·R·A·H·M·S KRYPTOR Compact Plus (ThermoFisher Scientific).

METHODS

We examined K2-EDTA plasma samples from 50 patients on B·R·A·H·M·S KRYPTOR Compact Plus, an automated immunoassay platform. QC materials were used to assess intra- and inter-precision of the assay. Lower limit of quantitation and interference studies were determined using pooled patient plasma.

RESULTS

The sFlt1 and PlGF assays demonstrated an analytical measuring range of 90-69,000 pg/mL and 11-7000 pg/mL, respectively (r2 > 0.99). Lower limit of quantitation (20% CV) was interpolated to be 35 pg/mL for sFlt1 and 10 pg/mL for PlGF. Total precision for both assay displayed CVs of <10%. Interference studies showed that both assays were not significantly affected by hemolysis up to an H-index of 1100 for sFlt1 and 300 for PlGF; L- and I-index of 800 and 80 respectively for both assays. The Passing-Bablok regression analysis for sFlt1/PlGF yielded an equation of y = 1.05x + 0.02, and the Bland Altman analysis showed an average bias of 0.84.

CONCLUSIONS

Plasma levels of sFlt1 and PlGF measured on the B·R·A·H·M·S KRYPTOR Compact Plus platform demonstrate excellent analytical performance and are acceptable as clinical grade assays.

Word lid van onze
facebookpagina

De meest complete database met geneeskrachtige kruiden, ondersteund door de wetenschap

  • Werkt in 55 talen
  • Kruidengeneesmiddelen gesteund door de wetenschap
  • Kruidenherkenning door beeld
  • Interactieve GPS-kaart - tag kruiden op locatie (binnenkort beschikbaar)
  • Lees wetenschappelijke publicaties met betrekking tot uw zoekopdracht
  • Zoek medicinale kruiden op hun effecten
  • Organiseer uw interesses en blijf op de hoogte van nieuwsonderzoek, klinische onderzoeken en patenten

Typ een symptoom of een ziekte en lees over kruiden die kunnen helpen, typ een kruid en zie ziekten en symptomen waartegen het wordt gebruikt.
* Alle informatie is gebaseerd op gepubliceerd wetenschappelijk onderzoek

Google Play badgeApp Store badge