Comparison of two three-dimensional gated SPECT methods with thallium in patients with large myocardial infarction.

Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
P Véra
R Koning
A Cribier
A Manrique

Sleutelwoorden

Abstract

BACKGROUND

Two different commercially available gated single photon emission computed tomography (GSPECT) methods were compared in a population of patients with a major myocardial infarction.

METHODS

Rest thallium GSPECT was performed with a 90-degree dual-detector camera, 4 hours after injection of thallium-201 (Tl-201; 185 MBq) in 43 patients (mean age, 62+/-12 years) with a large myocardial infarction (mean defect size, 33%+/-16%). End-diastolic volume (EDV), end-systolic volume (ESV), and left ventricular ejection fraction (LVEF) were calculated by using QGS (Cedars Sinai) and MultiDim (Sopha Medical Vision International, Buc, France). Images were reconstructed by using a 2.5 zoom and a Butterworth filter (order, 5; cut-off frequency, 0.20). LVEF was calculated in all patients by using equilibrium radionuclide angiocardiography (ERNA). EDV, ESV, and LVEF were also measured by using left ventriculography (LVG).

RESULTS

Compared with LVG, QGS underestimated LVEF by means of an underestimation of mean EDV. MultiDim overestimated EDV and ESV. GSPECT EDV and ESV overestimation was demonstrated by means of Bland-Altman analysis to increase with left ventricular volume size (P<.05). The difference between LVG and GSPECT volumes was demonstrated by means of regression analysis to be correlated with infarction size. This effect was particularly important with MultiDim (P<.0001).

CONCLUSIONS

In Tl-201 GSPECT, LVEF and volume measurements will vary according to the type of software used.

Word lid van onze
facebookpagina

De meest complete database met geneeskrachtige kruiden, ondersteund door de wetenschap

  • Werkt in 55 talen
  • Kruidengeneesmiddelen gesteund door de wetenschap
  • Kruidenherkenning door beeld
  • Interactieve GPS-kaart - tag kruiden op locatie (binnenkort beschikbaar)
  • Lees wetenschappelijke publicaties met betrekking tot uw zoekopdracht
  • Zoek medicinale kruiden op hun effecten
  • Organiseer uw interesses en blijf op de hoogte van nieuwsonderzoek, klinische onderzoeken en patenten

Typ een symptoom of een ziekte en lees over kruiden die kunnen helpen, typ een kruid en zie ziekten en symptomen waartegen het wordt gebruikt.
* Alle informatie is gebaseerd op gepubliceerd wetenschappelijk onderzoek

Google Play badgeApp Store badge