Biochemia Medica 2019-Jun
Haemolysis and lipemia interfere with resistin and myeloperoxidase BioVendor ELISA assays.
Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
Sleutelwoorden
Abstract
Materials and methods
Blood was taken from healthy volunteers into lithium heparin tubes. Plasma samples were spiked with Lipofundin® emulsion (B. Braun Melsungen AG, Germany) for lipemia interference testing. Haemolysed samples were obtained by drawing aliquots of heparinized blood through a 26 gauge needle. Index of haemolysis (H), lipemia (L) and triglyceride concentration were measured on Abbott Architect c8000. Haemoglobin concentration was measured on Sysmex XN-1000. Concentrations of RES and MPO in all samples were determined with RES and MPO ELISA kits (BioVendor, Czech Republic). All measurements were performed in triplicate. Biases from the native samples were calculated for both analytes and compared with an arbitrary value (e.g. ± 10%).