Dutch
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Medical Physics 2015-Jan

Markerless motion tracking of lung tumors using dual-energy fluoroscopy.

Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
Rakesh Patel
Joshua Panfil
Maria Campana
Alec M Block
Matthew M Harkenrider
Murat Surucu
John C Roeske

Sleutelwoorden

Abstract

OBJECTIVE

To evaluate the efficacy of dual-energy (DE) vs single-energy (SE) fluoroscopic imaging of lung tumors using a markerless template-based tracking algorithm.

METHODS

Ten representative patient breathing patterns were programmed into a Quasar™ motion phantom. The phantom was modified by affixing pork ribs to the surface, and a cedar insert with a small spherical volume was used to simulate lung and tumor, respectively. Sequential 60 kVp (6 mA) and 120 kVp (1.5 mA) fluoroscopic sequences were acquired. Frame-by-frame weighted logarithmic subtraction was performed resulting in a DE fluoroscopic sequence. A template-based algorithm was then used to track tumor motion throughout the DE and SE fluoroscopy sequences. Tracking coordinates were evaluated against ground-truth tumor locations. Fluoroscopic images were also acquired for two lung cancer patients, neither of which had implanted fiducials.

RESULTS

For phantom imaging, a total of 1925 frames were analyzed. The algorithm successfully tracked the target on 99.9% (1923/1925) of DE frames vs 90.7% (1745/1925) SE images (p < 0.01). The displacement between tracking coordinates and ground truth for the phantom was 1.4 mm ± 1.1 mm for DE vs 2.0 mm ± 1.3 mm for SE (p < 0.01). Images from two patients, one with a larger tumor and one with a smaller tumor, were also analyzed. For the patient with the larger tumor, the average displacement from physician defined ground truth was 1.2 mm ± 0.6 mm for DE vs 1.4 mm ± 0.7 mm for SE (p = 0.016). For the patient that presented with a smaller tumor, the average displacement from physician defined ground truth was 2.2 mm ± 1.0 mm for DE vs 3.2 mm ± 1.4 mm for SE (p < 0.01). Importantly, for this single patient with the smaller tumor, 15.6% of the SE frames had >5 mm displacements from the ground truth vs 0% for DE fluoroscopy.

CONCLUSIONS

This work indicates the potential for markerless tumor tracking utilizing DE fluoroscopy. With DE imaging, the algorithm showed improved detectability vs SE fluoroscopy and was able to accurately track the tumor in nearly all cases.

Word lid van onze
facebookpagina

De meest complete database met geneeskrachtige kruiden, ondersteund door de wetenschap

  • Werkt in 55 talen
  • Kruidengeneesmiddelen gesteund door de wetenschap
  • Kruidenherkenning door beeld
  • Interactieve GPS-kaart - tag kruiden op locatie (binnenkort beschikbaar)
  • Lees wetenschappelijke publicaties met betrekking tot uw zoekopdracht
  • Zoek medicinale kruiden op hun effecten
  • Organiseer uw interesses en blijf op de hoogte van nieuwsonderzoek, klinische onderzoeken en patenten

Typ een symptoom of een ziekte en lees over kruiden die kunnen helpen, typ een kruid en zie ziekten en symptomen waartegen het wordt gebruikt.
* Alle informatie is gebaseerd op gepubliceerd wetenschappelijk onderzoek

Google Play badgeApp Store badge