Journal of Biomedical Optics

Noninvasive method for measuring local hemoglobin oxygen saturation in tissue using wide gap second derivative near-infrared spectroscopy.

Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
Dean E Myers
LeAnn D Anderson
Roxanne P Seifert
Joseph P Ortner
Chris E Cooper
Greg J Beilman
John D Mowlem

Sleutelwoorden

Abstract

A simple continuous wave near-infrared algorithm for estimating local hemoglobin oxygen saturation in tissue (%StO2) is described using single depth attenuation measurements at 680, 720, 760, and 800 nm. Second derivative spectroscopy was used to reduce light scattering effects, chromophores with constant absorption, baseline/instrumentation drift, and movement artifacts. Unlike previous second derivative methods which focused primarily on measuring deoxyhemoglobin concentration; a wide 40 nm wavelength gap used for calculating second derivative attenuation significantly improved sensitivity to oxyhemoglobin absorption. Scaled second derivative attenuation at 720 nm was correlated to in vitro hemoglobin oxygen saturation to generate a %StO2 calibration curve. The calibration curve was insensitive to total hemoglobin, optical path length, and optical scattering. Measurement error due to normal levels of carboxyhemoglobin, methemoglobin, and water absorption were less than 10 %StO2 units. Severe methemoglobinemia or edema combined with low blood volume could cause StO2 errors to exceed 10 StO2 units. Both a broadband and commercial four-wavelength spectrometer (InSpectra) measured %StO2. The InSpectra tissue spectrometer readily detected limb ischemia on 26 human volunteers for hand, forearm, and leg muscles. A strong linear correlation, r2>0.93, between StO2 and microvascular %SO2 was observed for isolated animal hind limb, kidney, and heart.

Word lid van onze
facebookpagina

De meest complete database met geneeskrachtige kruiden, ondersteund door de wetenschap

  • Werkt in 55 talen
  • Kruidengeneesmiddelen gesteund door de wetenschap
  • Kruidenherkenning door beeld
  • Interactieve GPS-kaart - tag kruiden op locatie (binnenkort beschikbaar)
  • Lees wetenschappelijke publicaties met betrekking tot uw zoekopdracht
  • Zoek medicinale kruiden op hun effecten
  • Organiseer uw interesses en blijf op de hoogte van nieuwsonderzoek, klinische onderzoeken en patenten

Typ een symptoom of een ziekte en lees over kruiden die kunnen helpen, typ een kruid en zie ziekten en symptomen waartegen het wordt gebruikt.
* Alle informatie is gebaseerd op gepubliceerd wetenschappelijk onderzoek

Google Play badgeApp Store badge