Clinical Chemistry 2004-Feb

Quantification of free sialic acid in urine by HPLC-electrospray tandem mass spectrometry: a tool for the diagnosis of sialic acid storage disease.

Alleen geregistreerde gebruikers kunnen artikelen vertalen
Log in Schrijf in
De link wordt op het klembord opgeslagen
Fredoen Valianpour
Nicolaas G G M Abeling
Marinus Duran
Jan G M Huijmans
Willem Kulik

Sleutelwoorden

Abstract

BACKGROUND

Sialic acid storage diseases (SSDs) are severe autosomal recessive neurodegenerative disorders caused by a transport defect across the lysosomal membrane, which leads to accumulation of sialic acid in tissues, fibroblasts, and urine. Defective free sialic acid transport can be established by quantification of free sialic acid in urine.

METHODS

Urine sample size was adjusted to the equivalent of 100 nmol of creatinine. After addition of 2-keto-3-deoxy-d-glycero-d-galactonononic acid as internal standard, samples were diluted with water to an end volume of 250 microL. We used 10 microL for HPLC-tandem mass spectrometric analysis in the negative electrospray ionization mode, monitoring transitions m/z 308.3-->m/z 86.9 (sialic acid) and m/z 267.2-->m/z 86.9 (internal standard). The overall method was validated and studied for ion suppression, interfering compounds, and pH effects. Samples from controls (n = 72) and SSD patients (n = 3) were analyzed.

RESULTS

The limit of detection was 3 micromol/L. Intraassay imprecision (CV; n = 10) was 6%, 3%, and 2% at 30, 130, and 1000 mmol/mol creatinine, respectively; corresponding interassay CV (n = 10) were 5%, 5%, and 2%. Recovery was 109% (100-1000 mmol/mol creatinine). The mean (SD) [range] excretion rates (mmol/mol creatinine) were 31.3 (16.6) [0.7-56.9] at 0-1 year (n = 20), 21.2 (9.8) [6.3-38.3] at 1-3 years (n = 15), 14.4 (8.2) [1.7-32.9] at 3-10 years (n = 25), and 4.6 (2.6) [0-9.8] above age 10 years (n = 12). SSD patients 1.2, 3.9, and 12 years of age had concentrations of 111.5, 54.2, and 36.1 mmol/mol creatinine, respectively.

CONCLUSIONS

The HPLC-tandem MS method for free sialic acid in urine is more rapid, accurate, sensitive, selective, and robust than earlier methods and may serve as a candidate reference method for free sialic acid in diagnosis of SSD.

Word lid van onze
facebookpagina

De meest complete database met geneeskrachtige kruiden, ondersteund door de wetenschap

  • Werkt in 55 talen
  • Kruidengeneesmiddelen gesteund door de wetenschap
  • Kruidenherkenning door beeld
  • Interactieve GPS-kaart - tag kruiden op locatie (binnenkort beschikbaar)
  • Lees wetenschappelijke publicaties met betrekking tot uw zoekopdracht
  • Zoek medicinale kruiden op hun effecten
  • Organiseer uw interesses en blijf op de hoogte van nieuwsonderzoek, klinische onderzoeken en patenten

Typ een symptoom of een ziekte en lees over kruiden die kunnen helpen, typ een kruid en zie ziekten en symptomen waartegen het wordt gebruikt.
* Alle informatie is gebaseerd op gepubliceerd wetenschappelijk onderzoek

Google Play badgeApp Store badge