Hemodialysis international. International Symposium on Home Hemodialysis 2017-Aug

Comparison of stroke volume measurements during hemodialysis using bioimpedance cardiography and echocardiography.

Bare registrerte brukere kan oversette artikler
Logg inn Registrer deg
Koblingen er lagret på utklippstavlen
Michael J Germain
Jyovani Joubert
Daniel O'Grady
Brian H Nathanson
Yossi Chait
Nathan W Levin

Nøkkelord

Abstrakt

BACKGROUND

Fluid management remains a major challenge of hemodialysis (HD) care, with serious implications for morbidity and mortality. Intradialytic fluid management is typically guided by blood pressure, an indirect resultant of hemodynamics status. Direct measurements of hemodynamic parameters may improve cardiovascular outcomes by providing rational bases for intervention. We compare stroke volume (SV) measurements using a noninvasive, regional biompedance cardiography device (NiCaS) with Doppler echocardiography (Echo) in HD setting.

METHODS

Stroke volumes were simultaneously measured using the devices in 17 patients receiving maintenance HD. Measurements were made during 2 weekly HD treatments, and twice within each HD treatment during the first and last hour of each treatment, for a total of 64 SV measurements. Agreement between devices was assessed using linear regression, a Pearson's correlation coefficient, and a Bland-Altman plot all adjusted for repeated measures within patients.

RESULTS

Echo and NiCaS SV mean and 95% CIs were 58.0 (50.1, 65.8) and 56.7 (49.4, 64.0) mL, respectively. NiCaS SV correlated strongly with Echo SV during the first and last hours of treatments (r = 0.93, P < 0.001 and r = 0.92, P < 0.001, respectively). Linear regression of NiCaS on Echo showed a slope of 0.97, 95% CI (0.91, 1.02) which did not differ from 1, P = 0.20. A Bland-Altman plot and 4-Quadrant plot confirmed that the 2 methods produced comparable measurements.

CONCLUSIONS

NiCaS SV measurements are similar to and strongly correlated with Echo SV measurements. This suggests that noninvasive NiCaS technology may be a practical method for measuring SV during HD.

Bli med på
facebooksiden vår

Den mest komplette databasen med medisinske urter støttet av vitenskap

  • Fungerer på 55 språk
  • Urtekurer støttet av vitenskap
  • Urtegjenkjenning etter bilde
  • Interaktivt GPS-kart - merk urter på stedet (kommer snart)
  • Les vitenskapelige publikasjoner relatert til søket ditt
  • Søk medisinske urter etter deres effekter
  • Organiser dine interesser og hold deg oppdatert med nyheter, kliniske studier og patenter

Skriv inn et symptom eller en sykdom og les om urter som kan hjelpe, skriv en urt og se sykdommer og symptomer den brukes mot.
* All informasjon er basert på publisert vitenskapelig forskning

Google Play badgeApp Store badge