European Journal of Radiology 2012-Dec

Triple-rule-out dual-source CT angiography of patients with acute chest pain: dose reduction potential of 100 kV scanning.

Bare registrerte brukere kan oversette artikler
Logg inn Registrer deg
Koblingen er lagret på utklippstavlen
Radko Krissak
Thomas Henzler
Anne Prechel
Miriam Reichert
Joachim Gruettner
Tim Sueselbeck
Stefan O Schoenberg
Christian Fink

Nøkkelord

Abstrakt

OBJECTIVE

To investigate the dose reduction potential of low kV triple-rule-out dual-source CT angiography (TRO-CTA) in non-obese (BMI ≤ 25 kg/m(2)) patients with acute chest pain.

METHODS

Sixty consecutive patients were randomly assigned to two different retrospectively ECG-gated TRO-CTA protocols in this prospective trial: Thirty patients were examined with a 120-kV standard protocol (320 reference mAs with automatic tube current modulation, automatically adapted pitch and ECG-pulsing) and served as the control group (group 1), an otherwise identical 100 kV protocol was used in the other thirty patients (group 2) for a radiation dose reduction. Subjective image quality was assessed on a 5 point scale (1: excellent, 5: non-diagnostic) by two blinded observers. Quantitative image analysis assessed vascular attenuation, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in different vascular segments. The effective dose was calculated from the dose length product (DLP) using a conversion coefficient of 0.017 mSvmGy(-1)cm(-1).

RESULTS

There was no significant difference of age, BMI, heart rate, pitch or scan length between both patient groups. Subjective image quality was rated similar in both groups (group 1: 1.2 ± 0.4, group 2: average score=1.3 ± 0.5). Vessel attenuation was significantly higher in group 2 than in group 1 (ascending aorta: 456 ± 83 HU vs. 370 ± 78 HU, p<0.001; pulmonary artery: 468 ± 118 HU vs. 411 ± 91 HU, p=0.03; left coronary artery: 437 ± 110 HU vs. 348 ± 89 HU, p<0.001), however, there was no significant difference in SNR (13.2 ± 7.6 vs. 14.5 ± 7.5, p=0.49) or CNR (13.8 ± 6.6 vs. 15.9 ± 7.7, p=0.25). The effective radiation dose of the 100 kV protocol was significantly lower (9.6 ± 3.2 mSv vs. 18.1 ± 9.4 mSv, p<0.0001).

CONCLUSIONS

TRO-CTA with 100 kV is feasible in non-obese patients and results in diagnostic image quality and significantly reduced radiation dose.

Bli med på
facebooksiden vår

Den mest komplette databasen med medisinske urter støttet av vitenskap

  • Fungerer på 55 språk
  • Urtekurer støttet av vitenskap
  • Urtegjenkjenning etter bilde
  • Interaktivt GPS-kart - merk urter på stedet (kommer snart)
  • Les vitenskapelige publikasjoner relatert til søket ditt
  • Søk medisinske urter etter deres effekter
  • Organiser dine interesser og hold deg oppdatert med nyheter, kliniske studier og patenter

Skriv inn et symptom eller en sykdom og les om urter som kan hjelpe, skriv en urt og se sykdommer og symptomer den brukes mot.
* All informasjon er basert på publisert vitenskapelig forskning

Google Play badgeApp Store badge