Journal of Interventional Cardiology 2018-Jun

Cardiovascular magnetic resonance assessment of 1st generation CoreValve and 2nd generation Lotus valves.

Tylko zarejestrowani użytkownicy mogą tłumaczyć artykuły
Zaloguj się Zarejestruj się
Link zostanie zapisany w schowku
Tarique Al Musa
Akhlaque Uddin
Laura E Dobson
Peter P Swoboda
Pankaj Garg
James R J Foley
Christopher Malkin
Sven Plein
Daniel J Blackman
John P Greenwood

Słowa kluczowe

Abstrakcyjny

OBJECTIVE

We sought to compare using serial CMR, the quantity of AR and associated valve hemodynamics, following the first-generation CoreValve (Medtronic, Minneapolis, MN) and the second-generation Lotus valve (Boston Scientific, Natick, MA).

BACKGROUND

Aortic regurgitation (AR) following Transcatheter Aortic Valve Replacement (TAVR) confers a worse prognosis and can be accurately quantified using cardiovascular magnetic resonance (CMR). Second generation valves have been specifically designed to reduce paravalvular AR and improve clinical outcomes.

METHODS

Fifty-one patients (79.0 ± 7.7 years, 57% male) were recruited and imaged at three time points: immediately pre- and post-TAVR, and at 6 months.

RESULTS

CMR-derived AR fraction immediately post-TAVR was greater in the CoreValve compared to Lotus group (11.7 ± 8.4 vs. 4.3 ± 3.4%, P = 0.001), as was the frequency of ≥moderate AR (9/24 (37.5%) versus 0/27, P < 0.001). However, at 6 months AR fraction had improved significantly in the CoreValve group such that the two valve designs were comparable (6.4 ± 5.0 vs 5.6 ± 5.3%, P = 0.623), with no patient in either group having ≥moderate AR. The residual peak pressure gradient immediately following TAVR was significantly lower with CoreValve compared to Lotus (14.1 ± 5.6 vs 25.4 ± 11.6 mmHg, P = 0.001), but again by 6 months the two valve designs were comparable (16.5 ± 9.4 vs 19.7 ± 10.5 mmHg, P = 0.332). There was no difference in the degree of LV reverse remodeling between the two valves at 6 months.

CONCLUSIONS

Immediately post-TAVR, there was significantly less AR but a higher residual peak pressure gradient with the Lotus valve compared to CoreValve. However, at 6 months both devices had comparable valve hemodynamics and LV reverse remodeling.

Dołącz do naszej strony
na Facebooku

Najbardziej kompletna baza danych ziół leczniczych poparta naukowo

  • Działa w 55 językach
  • Ziołowe leki poparte nauką
  • Rozpoznawanie ziół na podstawie obrazu
  • Interaktywna mapa GPS - oznacz zioła na miejscu (wkrótce)
  • Przeczytaj publikacje naukowe związane z Twoim wyszukiwaniem
  • Szukaj ziół leczniczych po ich działaniu
  • Uporządkuj swoje zainteresowania i bądź na bieżąco z nowościami, badaniami klinicznymi i patentami

Wpisz objaw lub chorobę i przeczytaj o ziołach, które mogą pomóc, wpisz zioło i zobacz choroby i objawy, na które są stosowane.
* Wszystkie informacje oparte są na opublikowanych badaniach naukowych

Google Play badgeApp Store badge