CardioVascular and Interventional Radiology

Hyperthermia in the treatment of cholangiocarcinoma: development and testing of an endobiliary microwave device.

Tylko zarejestrowani użytkownicy mogą tłumaczyć artykuły
Zaloguj się Zarejestruj się
Link zostanie zapisany w schowku
Sara Mantero
Iginio Longo
Gianfranco Beniamino Fiore
Aldo Severini

Słowa kluczowe

Abstrakcyjny

OBJECTIVE

The aim of this study was to create, perfect and test a hyperthermia balloon catheter for local treatment of cholangiocarcinoma. The device should induce hyperthermia in tumor tissue by acting locally in the bile duct lumen in contact with the tumor-infiltrated duct wall. In addition, it should exert tissue compression to cause an appreciable reduction in tumor microvasculature flow, thus improving thermal performance.

METHODS

According to the design specifications, the working temperature range should allow the device to be used for hyperthermia therapy either in combination with radiation and/or chemotherapy (radio/chemo-sensitizing effect at 42-45 degrees C) or alone (induction of coagulation necrosis above 65 degrees C). The balloon serves as an anchoring system during treatment and as a functional element to induce tissue compression. In vitro mechanical evaluation of the pressure/volume relationship, with the balloon inside rigid walled conduits, was performed. The heating shape around the catheter tip was determined by egg-white heating tests (coagulation at about 65 degrees C). Moreover, heating tests were carried out with explanted pig liver parenchyma. The temperature profile over time at different depths from the catheter axis was traced.

RESULTS

Three prototypes were manufactured. Mechanical functional tests showed that a 14 Fr deflated diameter balloon was suitable for bile duct diameters up to 11 mm. Thermal egg-white tests produced 4 cm long, 3 cm in diameter ellipsoidal heating figures in 30 min. In the biological tissue tests a coagulated area of similar geometry and comparable volume was produced.

CONCLUSIONS

The results of tests confirm the device's usefulness and versatility.

Dołącz do naszej strony
na Facebooku

Najbardziej kompletna baza danych ziół leczniczych poparta naukowo

  • Działa w 55 językach
  • Ziołowe leki poparte nauką
  • Rozpoznawanie ziół na podstawie obrazu
  • Interaktywna mapa GPS - oznacz zioła na miejscu (wkrótce)
  • Przeczytaj publikacje naukowe związane z Twoim wyszukiwaniem
  • Szukaj ziół leczniczych po ich działaniu
  • Uporządkuj swoje zainteresowania i bądź na bieżąco z nowościami, badaniami klinicznymi i patentami

Wpisz objaw lub chorobę i przeczytaj o ziołach, które mogą pomóc, wpisz zioło i zobacz choroby i objawy, na które są stosowane.
* Wszystkie informacje oparte są na opublikowanych badaniach naukowych

Google Play badgeApp Store badge