World Neurosurgery 2016-Apr

Implant Site Infection and Bone Flap Osteomyelitis Associated with the NeuroPace Responsive Neurostimulation System.

Tylko zarejestrowani użytkownicy mogą tłumaczyć artykuły
Zaloguj się Zarejestruj się
Link zostanie zapisany w schowku
Zhikui Wei
Chad R Gordon
Gregory K Bergey
Justin M Sacks
William S Anderson

Słowa kluczowe

Abstrakcyjny

BACKGROUND

The NeuroPace RNS System is a method recently approved by the U.S. Food and Drug Administration for closed-loop direct brain stimulation in selected patients with drug-resistant partial seizures. The long-term risks of implant site infection and accompanying bone flap osteomyelitis associated with responsive neurostimulation (RNS) devices have not been fully appreciated.

METHODS

We report 3 cases of refractory partial epilepsy that were treated with RNS therapy in conjunction with antiepileptic drugs. Patients underwent invasive epilepsy monitoring and resection of seizure foci. All patients continued to have debilitating partial seizures and underwent implantation of the RNS device, which resulted in various degrees of symptomatic relief. On average, the battery of the implantable pulse generator was replaced every 2 years. All 3 patients developed implant site infection and bone flap osteomyelitis with multiple implantable pulse generator replacements, and the RNS devices were removed. Bone flaps were removed in 2 patients because of significant osteomyelitis and were reconstructed in a delayed fashion with customized cranial implants. No patient had evidence of meningitis or cerebritis. The patients were treated via a multidisciplinary approach, and all patients recovered well with satisfactory wound healing and seizure control.

CONCLUSIONS

Implant site infection and bone flap osteomyelitis are significant adverse events associated with the RNS device. The incidence of infection in this series (10%) is comparable to the incidence reported in the long-term trial. The infection risk is mainly associated with reoperations and increases with multiple implantable pulse generator replacements. The RNS device may benefit from reducing technical risk factors that are associated with postoperative bone and soft tissue infections.

Dołącz do naszej strony
na Facebooku

Najbardziej kompletna baza danych ziół leczniczych poparta naukowo

  • Działa w 55 językach
  • Ziołowe leki poparte nauką
  • Rozpoznawanie ziół na podstawie obrazu
  • Interaktywna mapa GPS - oznacz zioła na miejscu (wkrótce)
  • Przeczytaj publikacje naukowe związane z Twoim wyszukiwaniem
  • Szukaj ziół leczniczych po ich działaniu
  • Uporządkuj swoje zainteresowania i bądź na bieżąco z nowościami, badaniami klinicznymi i patentami

Wpisz objaw lub chorobę i przeczytaj o ziołach, które mogą pomóc, wpisz zioło i zobacz choroby i objawy, na które są stosowane.
* Wszystkie informacje oparte są na opublikowanych badaniach naukowych

Google Play badgeApp Store badge