Polish
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
BMC Pediatrics 2013-Sep

Validation of a transcutaneous bilirubin meter in Mongolian neonates: comparison with total serum bilirubin.

Tylko zarejestrowani użytkownicy mogą tłumaczyć artykuły
Zaloguj się Zarejestruj się
Link zostanie zapisany w schowku
Moe Akahira-Azuma
Naohiro Yonemoto
Battsengel Ganzorig
Rintaro Mori
Shinichi Hosokawa
Takeji Matsushita
Bayasgalantai Bavuusuren
Enkhtur Shonkhuuz

Słowa kluczowe

Abstrakcyjny

BACKGROUND

Neonatal hyperbilirubinemia, especially kernicterus, can be prevented by screening for neonatal jaundice. The transcutaneous bilirubin (TcB) meter is a non-invasive medical device for screening neonates. The study aimed to investigate the validity of a TcB meter in a resource-limited setting such as Mongolia.

METHODS

Term and late preterm neonates from the National Center for Maternal and Child Health of Ulaanbaatar in Mongolia who met the inclusion criteria (gestational age ≥35 weeks, birth weight ≥2000 g, postnatal age ≤ 1 month) were enrolled in the study. We used a TcB meter, JM-103 to screen for neonatal jaundice. TcB measurements at the infant's forehead and midsternum were performed within 3 h of obtaining samples for total serum bilirubin (TSB) measurement. We analyzed the correlation between TcB measurements and TSB measurements to validate the meter.

RESULTS

A total of 47 term and six late preterm neonates were included in the study. TcB measured by the meter at both the forehead and the midsternum showed a strong correlation with TSB measured in the laboratory. The correlation equations were TSB = 1.409+0.8655 × TcB (R2=0.78871) at the forehead, and TSB = 0.7555+0.8974 × TcB (R2=0.78488) at the midsternum. Bland-Altman plots and the Bradley-Blackwood test showed no significant differences between the two methods at all measured ranges of bilirubin. The mean areas under the curves of TcB at the forehead and midsternum at three TSB levels (>10 mg/dL, >13 mg/dL, >15 mg/dL) of TcB were greater than 0.9, and all had high sensitivity and specificity.

CONCLUSIONS

This study established the validity of the JM-103 meter as a screening tool for neonatal jaundice in term and late preterm infants in Mongolia. Future studies are needed, including the establishment of a TcB hour-specific nomogram, for more effective clinical practice to prevent severe hyperbilirubinemia.

Dołącz do naszej strony
na Facebooku

Najbardziej kompletna baza danych ziół leczniczych poparta naukowo

  • Działa w 55 językach
  • Ziołowe leki poparte nauką
  • Rozpoznawanie ziół na podstawie obrazu
  • Interaktywna mapa GPS - oznacz zioła na miejscu (wkrótce)
  • Przeczytaj publikacje naukowe związane z Twoim wyszukiwaniem
  • Szukaj ziół leczniczych po ich działaniu
  • Uporządkuj swoje zainteresowania i bądź na bieżąco z nowościami, badaniami klinicznymi i patentami

Wpisz objaw lub chorobę i przeczytaj o ziołach, które mogą pomóc, wpisz zioło i zobacz choroby i objawy, na które są stosowane.
* Wszystkie informacje oparte są na opublikowanych badaniach naukowych

Google Play badgeApp Store badge