Monitoring Noninvasive Ventilation in Patients with Obesity Hypoventilation Syndrome: Comparison between Ventilator Built-in Software and Respiratory Polygraphy.

Apenas usuários registrados podem traduzir artigos
Entrar Inscrever-se
O link é salvo na área de transferência
Ramon Fernandez Alvarez
Claudio Rabec
Gemma Rubinos Cuadrado
Juan Alejandro Cascon Hernandez
Patricia Rodriguez
Marjolaine Georges
Pere Casan

Palavras-chave

Resumo

BACKGROUND

Polygraphy (PG) remains the standard method of assessing noninvasive ventilation (NIV) effectiveness. Built-in software (BIS) of recent NIV equipment provides estimates of some ventilator parameters, but their usefulness is unclear.

OBJECTIVE

To assess the reliability of BIS compared with PG in a cohort of obesity hypoventilation syndrome (OHS) patients on NIV.

METHODS

Thirty stable OHS patients on NIV were evaluated in an outpatient setting with simultaneous PG and BIS recordings. The automated apnea-hypopnea event index (EIAUT) provided by Rescan and manual scoring based on available traces obtained from the software (EIBIS) were compared with manual PG scoring (EIPG). Each manual scoring was separately performed by 2 trained operators. Agreement between the 2 operators was assessed using the kappa coefficient. Pearson correlation and Bland-Altman plots were used to evaluate agreement between EIAUT, EIBIS, and EIPG.

RESULTS

Twenty-six cases were valid for analysis (age ±61 years, 17 men). All patients were ventilated in the spontaneous/timed mode (mean inspiratory positive airway pressure 17 ± 3 cm H2O, mean expiratory positive airway pressure 10 ± 3 cm H2O). Cohen's kappa agreement between the operators was 0.7 for EIBIS and 0.84 for EIPG. EIBIS showed good correlation with EIPG (r2 = 0.79 p < 0.001), better than scoring provided by the automated analysis (r2 = 0.71, p < 0.006 for EIAUT vs. EIPG).

CONCLUSIONS

In stable OHS patients on NIV, unattended home-based monitoring using Rescan is reproducible and reliable to assess quality of ventilation when compared with PG. In addition, manual scoring of events using data obtained with this device is more consistent than software-based automated analysis.

Junte-se à nossa
página do facebook

O mais completo banco de dados de ervas medicinais apoiado pela ciência

  • Funciona em 55 idiomas
  • Curas herbais apoiadas pela ciência
  • Reconhecimento de ervas por imagem
  • Mapa GPS interativo - marcar ervas no local (em breve)
  • Leia publicações científicas relacionadas à sua pesquisa
  • Pesquise ervas medicinais por seus efeitos
  • Organize seus interesses e mantenha-se atualizado com as notícias de pesquisa, testes clínicos e patentes

Digite um sintoma ou doença e leia sobre ervas que podem ajudar, digite uma erva e veja as doenças e sintomas contra os quais ela é usada.
* Todas as informações são baseadas em pesquisas científicas publicadas

Google Play badgeApp Store badge