Portuguese
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Anaesthesia and Intensive Care 1993-Oct

The Australian Incident Monitoring Study. The capnograph: applications and limitations--an analysis of 2000 incident reports.

Apenas usuários registrados podem traduzir artigos
Entrar Inscrever-se
O link é salvo na área de transferência
J A Williamson
R K Webb
J Cockings
C Morgan

Palavras-chave

Resumo

The first 2000 incidents reported to the Australian Incident Monitoring Study were analysed with respect to the role of the capnograph. One hundred and fifty-seven (8%) were first detected by a capnograph and there were a further 18 (1%) in which capnography was contributory. Of the 1256 incidents which occurred in association with general anaesthesia 48% were "human detected" and 52% "monitor detected". The capnograph was ranked second and detected 24% of these monitor detected incidents; this figure would have been nearly 30% if a correctly checked, calibrated capnograph had always been used. The capnograph is a "front-line" monitor for oesophageal intubation, failure of ventilation, anaesthetic circuit faults, gas embolism, sudden circulatory collapse and malignant hyperthermia. It is a valuable "back-up" monitor when other monitors (e.g. low pressure alarm, pulse oximeter) are not in use, are being used incorrectly or fail. Such situations, in order of frequency of detection were: circuit-leak, overpressure of the breathing circuit, bronchospasm, leak of ventilator-driving-gas into the patient circuit, aspiration and/or regurgitation and hypoventilation. There were 20 reports of "failure", over two-thirds of which would not have occurred with appropriate checking and calibration. Seven were due to gas sampling problems and 6 to apnoea alarm failure. Two circuit leaks and 2 faulty unidirectional valves were not detected; on 3 occasions problems occurred due to power failure, calibration problems, or misinterpretation of an alarm.(ABSTRACT TRUNCATED AT 250 WORDS)

Junte-se à nossa
página do facebook

O mais completo banco de dados de ervas medicinais apoiado pela ciência

  • Funciona em 55 idiomas
  • Curas herbais apoiadas pela ciência
  • Reconhecimento de ervas por imagem
  • Mapa GPS interativo - marcar ervas no local (em breve)
  • Leia publicações científicas relacionadas à sua pesquisa
  • Pesquise ervas medicinais por seus efeitos
  • Organize seus interesses e mantenha-se atualizado com as notícias de pesquisa, testes clínicos e patentes

Digite um sintoma ou doença e leia sobre ervas que podem ajudar, digite uma erva e veja as doenças e sintomas contra os quais ela é usada.
* Todas as informações são baseadas em pesquisas científicas publicadas

Google Play badgeApp Store badge