Journal of Chromatography A 2017-Sep

Fully international system of units-traceable glycated hemoglobin quantification using two stages of isotope-dilution high-performance liquid chromatography-tandem mass spectrometry.

Только зарегистрированные пользователи могут переводить статьи
Войти Зарегистрироваться
Ссылка сохраняется в буфер обмена
Thi Thanh Huong Tran
Jinsook Lim
Juok Kim
Ha-Jeong Kwon
Gye Cheol Kwon
Ji-Seon Jeong

Ключевые слова

абстрактный

Glycated hemoglobin (HbA1c), defined as hemoglobin (Hb) molecules having a stable adduct of glucose on the N-terminal of the β-chains, has been endorsed as a diagnostic tool for diabetes mellitus and a prediction indicator for the development of diabetes complications. Here we describe an accurate procedure using two stages of isotope dilution-ultra performance liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) for the quantification of HbA1c that provides full traceability to International System of Units (SI). First, synthetic peptides representing specific markers of HbA1c (G-hexa) and hemoglobin A0 (Hexa) were certified by amino acid analysis via acid hydrolysis as reference materials (RMs) for the next step. For this peptide certification, three amino acids (proline, valine, and leucine) were determined by hydrolysis with 10M hydrochloric acid at 130°C for 48h followed by ID-LC-MS/MS. Then, HbA1c content in blood was quantified with the ratio of specific proteolytic peptides from HbA1c and HbA0 via enzyme digestion using ID-LC-MS/MS with the certified peptides as RMs and isotope-labeled peptides as internal standards. Results demonstrate complete traceability to SI-units throughout this procedure. Reliability was confirmed through comparative studies with commercially available RMs for HbA1c, and other routine HbA1c diagnostic methods as well. Following full method validation, we applied this procedure to the certification of candidate hemolysate-certified RMs for HbA1c content, as well as 52 real clinical samples. All of the results showed the suitability of this method to act as a primary reference measurement procedure for HbA1c in complex biological samples.

Присоединяйтесь к нашей
странице facebook

Самая полная база данных о лекарственных травах, подтвержденная наукой

  • Работает на 55 языках
  • Травяные лекарства, подтвержденные наукой
  • Распознавание трав по изображению
  • Интерактивная карта GPS - отметьте травы на месте (скоро)
  • Прочтите научные публикации, связанные с вашим поиском
  • Ищите лекарственные травы по их действию
  • Организуйте свои интересы и будьте в курсе новостей исследований, клинических испытаний и патентов

Введите симптом или заболевание и прочтите о травах, которые могут помочь, введите лекарство и узнайте о болезнях и симптомах, против которых оно применяется.
* Вся информация основана на опубликованных научных исследованиях.

Google Play badgeApp Store badge