Slovak
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
International Journal of Hyperthermia 2004-Aug

A feasibility study of interstitial hyperthermia plus external beam radiotherapy in glioblastoma multiforme using the Multi ELectrode Current Source (MECS) system.

Články môžu prekladať iba registrovaní používatelia
Prihlásiť Registrácia
Odkaz sa uloží do schránky
M C C M Hulshof
B W Raaymakers
J J W Lagendijk
R W Koot
H Crezee
L J A Stalpers
D González González

Kľúčové slová

Abstrakt

OBJECTIVE

Thermoradiotherapy has been shown in several randomized trials to increase local control compared to radiotherapy alone. The first randomized study of interstitial hyperthermia in glioblastoma multiforme showed a survival benefit for hyperthermia, though small. Improvement of the heating technique could lead to improved results. The purpose of this feasibility study is to present the clinical and thermal data of application of an improved interstitial hyperthermia system.

METHODS

Six patients with a glioblastoma multiforme were treated with interstitial hyperthermia using the Multi Electrode Current Source Interstitial Hyperthermia (MECS-IHT) system. The MECS-IHT system has the capability of spatial monitoring of temperature and individually steering of heating electrodes. Three sessions were given aiming at a steady state temperature of 42 degrees C for 1 h, with an interval of 3-4 days, during an external irradiation scheme of 60 Gy in 6 weeks. Hyperthermia was delivered with a mean of 10 catheters, 18 heating electrodes and 38 thermal probes per patient.

RESULTS

Sub-optimal temperatures were encountered in the first two patients leading to adjustments in technique thereafter with subsequent improvement of thermal data. With a catheter spacing of 11-12 mm, measurements yielded a mean T(90), T(50) and T(10) of 39.9, 43.7 and 45.2 degrees C, respectively, over three sessions in the last patient. The power per electrode to reach this temperature distribution varied from 25-100% of full power in each of the last four patients. Thermal data were reproducible over the three sessions. Acute toxicity was minimal.

CONCLUSIONS

Despite the spatial steering capabilities of the MECS-IHT system, a large temperature heterogeneity was encountered. The heterogeneity was the reason to limit the catheter spacing to 11-12 mm, thus making only small tumour volumes feasible for interstitial heating.

Pripojte sa k našej
facebookovej stránke

Najkompletnejšia databáza liečivých bylín podporovaná vedou

  • Pracuje v 55 jazykoch
  • Bylinné lieky podporené vedou
  • Rozpoznávanie bylín podľa obrázka
  • Interaktívna GPS mapa - označte byliny na mieste (už čoskoro)
  • Prečítajte si vedecké publikácie týkajúce sa vášho hľadania
  • Vyhľadajte liečivé byliny podľa ich účinkov
  • Usporiadajte svoje záujmy a držte krok s novinkami, klinickými skúškami a patentmi

Zadajte príznak alebo chorobu a prečítajte si o bylinách, ktoré by vám mohli pomôcť, napíšte bylinu a pozrite sa na choroby a príznaky, proti ktorým sa používa.
* Všetky informácie sú založené na publikovanom vedeckom výskume

Google Play badgeApp Store badge