Slovak
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Journal of Veterinary Emergency and Critical Care 2016-Sep

Validation of the Accutrend Plus point-of-care triglyceride analyzer in horses, ponies, and donkeys.

Články môžu prekladať iba registrovaní používatelia
Prihlásiť Registrácia
Odkaz sa uloží do schránky
Michaela C Pongratz
Hannah K Junge
Barbara Riond
Colin C Schwarzwald

Kľúčové slová

Abstrakt

OBJECTIVE

To assess the accuracy and reliability of a point-of-care (POC) triglyceride analyzer and to establish reference intervals for blood ([TRIG]POC/WB ) and plasma triglyceride concentrations ([TRIG]POC/PL ) in horses, ponies, and donkeys.

METHODS

Prospective study.

METHODS

University teaching hospital.

METHODS

120 adult healthy equids (78 horses and ponies, 42 donkeys) and 79 equids suffering from hypertriglyceridemia (73 horses and ponies, 6 donkeys).

METHODS

None.

RESULTS

[TRIG]POC/WB and [TRIG]POC/PL were measured using a POC analyzer and plasma triglyceride concentrations were measured using a standard laboratory assay ([TRIG]LAB/PL ). Reference intervals were determined. Test accuracy was assessed by Bland-Altman comparison of POC measurements with standard laboratory measurements and by evaluating linearity of dilutional series. Test reliability was assessed by repeated serial measurements. [TRIG]POC/WB and [TRIG]POC/PL were below the analytic range of the POC assay (<0.8 mmol/L [<70 mg/dL]) in healthy horses and ponies, whereas the reference intervals were 0.82-3.14 mmol/L (73-278 mg/dL) and 0.87-3.02 mmol/L (77-267 mg/dL), respectively, in donkeys. The POC analyzer systematically overestimated triglyceride concentrations when compared to a standard laboratory assay. The difference between [TRIG]POC/WB and [TRIG]POC/PL was small and clinically negligible. The coefficient of variation of repeated measures performed on the POC analyzer was below 10% for [TRIG]POC/WB and [TRIG]POC/PL , both in horses and donkeys and at all concentration ranges.

CONCLUSIONS

The POC analyzer allows accurate and reliable measurement of [TRIG]POC/WB and [TRIG]POC/PL in horses, ponies, and donkeys in clinical settings. Assay-specific reference intervals should be determined for diagnosis and clinical monitoring of hypertriglyceridemia in equids.

Pripojte sa k našej
facebookovej stránke

Najkompletnejšia databáza liečivých bylín podporovaná vedou

  • Pracuje v 55 jazykoch
  • Bylinné lieky podporené vedou
  • Rozpoznávanie bylín podľa obrázka
  • Interaktívna GPS mapa - označte byliny na mieste (už čoskoro)
  • Prečítajte si vedecké publikácie týkajúce sa vášho hľadania
  • Vyhľadajte liečivé byliny podľa ich účinkov
  • Usporiadajte svoje záujmy a držte krok s novinkami, klinickými skúškami a patentmi

Zadajte príznak alebo chorobu a prečítajte si o bylinách, ktoré by vám mohli pomôcť, napíšte bylinu a pozrite sa na choroby a príznaky, proti ktorým sa používa.
* Všetky informácie sú založené na publikovanom vedeckom výskume

Google Play badgeApp Store badge